Incoming-Outgoing Devices Coating Acceptance without Bakeout Pumping Speed Acceptance with Bakeout Leak Check Safety im Labor
  Incoming-Outgoing Devices  Not logged in ELOG logo
Message ID: 75     Entry time: Tue May 19 12:03:40 2020
Author: Phe 
Category: Problem 
Subject: Turbo Pump from Thomas Seiber 
Expected:  
Current Status: On schedule 
New Expected Date:  
Date Received: KW19 
Status: DONE 
Visual Inspection: Not Needed 
Test Started: 19.05.2020 
Setup: Customised 
Test Ended: 20.05.2020 
Result: Refused 
Outgoing Date: 27.05.2020 
Taken by: Student from T.Sieber 
Die Pumpe hat den Fehlercode 088 und hat eine Störung des Axiallagersensors.(checked by Graziano)
Attachment 1: pump_Sieber.jpg  98 kB  Uploaded Wed May 20 14:34:22 2020  | Hide | Hide all
pump_Sieber.jpg
ELOG V3.1.5-fc6679b