| Incoming-Outgoing Devices |
Not logged in |
 |
|
|
Message ID: 90
Entry time: Mon Sep 21 11:31:47 2020
|
| Author: |
Cristina |
| Category: |
test without bakeout |
| Subject: |
HEBT BPM |
| Expected: |
KW39 |
| Current Status: |
On schedule |
| New Expected Date: |
|
| Date Received: |
28.09.20 |
| Status: |
on-going |
| Visual Inspection: |
Accepted |
| Test Started: |
05.10.20. Test started again: 08.12.20 |
| Setup: |
Mobile 3 |
| Test Ended: |
|
| Result: |
Refused |
| Outgoing Date: |
|
| Taken by: |
|
|
|
| Since 19. Okt we wait for the decision of the diagnostic group on how(who has to clean the BPM. Since then the test is stopped.
Meeting on the 2.12.20: The japanise company will clean a BPM and they will send us to be tested.
Email from Dorn on the 07.12.20: we have to test the other 3 BPM, and later they will take care of the cleaning of the refused BPM.
On the 7.12 we tested another BPM: dirty. We sent on the 14.12 an email to Dorn and Kowina, showing that the ceramic are dirty and that they make dirty also our reference chamber. Measurement stopped again. |