Message ID: 90     Entry time: Mon Sep 21 11:31:47 2020
Author: Cristina 
Category: test without bakeout 
Subject: HEBT BPM 
Expected: KW39 
Current Status: On schedule 
New Expected Date:  
Date Received: 28.09.20 
Status: on-going 
Visual Inspection: Accepted 
Test Started: 05.10.20. Test started again: 08.12.20 
Setup: Mobile 3 
Test Ended:  
Result: Refused 
Outgoing Date:  
Taken by:  
Since 19. Okt we wait for the decision of the diagnostic group on how(who has to clean the BPM. Since then the test is stopped. Meeting on the 2.12.20: The japanise company will clean a BPM and they will send us to be tested. Email from Dorn on the 07.12.20: we have to test the other 3 BPM, and later they will take care of the cleaning of the refused BPM. On the 7.12 we tested another BPM: dirty. We sent on the 14.12 an email to Dorn and Kowina, showing that the ceramic are dirty and that they make dirty also our reference chamber. Measurement stopped again.