HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018
  Dielectronic recombination-assisted laser spectroscopy  Not logged in ELOG logo
Message ID: 100     Entry time: Tue May 31 19:32:56 2022
Author: Carsten, Wilfried, Max 
Category: Calibration 
Subject: electron cooler calibration voltage - frequency 
SetVoltage (DeltaU) MeasVoltage freq Current
+1950V 225247V 245.06725 50 mA
+1950V 225255.4V 245.0595 100 mA
+1960V 225266.4V 245.062125 100 mA
+1970V 225276.8V 245.064185 100 mA
+1980V 225286.6V 245.066 100 mA
+1990V 225296.1V 245.06775 100 mA
+2000V 225305.1V 245.0695 100 mA
+1950V 225255.5V 245.06 100 mA
+1940V 225246.0V 245.058125 100 mA
+1930V 225236.0V 245.05625 100 mA
+1920V 225226.0V 245.0545 100 mA
+1910V 225216.7V 245.052625 100 mA
+1900V 225205.8V 245.050375 100 mA
+1850V 225156.5V 245.041 100 mA
+1800V 225106.6V 245.031375 100 mA
+1950V 225256.1V 245.060125 100 mA
+2000V 225306.0V 245.06975 100 mA
+2050V 225355.8V 245.079375 100 mA
+2100V 225406.0V 245.0890 100 mA
+1950V 225256.2V 245.060125 100 mA
Change electron current and note down High Vltage and revolution frequency
NTCAP was restarted at the beginning and stopped at the end of the measuring period.
The NTCAP folder is 05-31_17-16-49 (system time 19:11)