HOME
ESR
EXPERIMENTS
LABS etc
|
E121
E127
E132
E125
E143
laser_cooling_2021
E142
Ê128
E146
E0052
E0018
|
Dielectronic recombination-assisted laser spectroscopy |
Not logged in |
|
|
Message ID: 100
Entry time: Tue May 31 19:32:56 2022
|
Author: |
Carsten, Wilfried, Max |
Category: |
Calibration |
Subject: |
electron cooler calibration voltage - frequency |
|
|
SetVoltage (DeltaU) | MeasVoltage | freq | Current
|
+1950V | 225247V | 245.06725 | 50 mA
|
+1950V | 225255.4V | 245.0595 | 100 mA
|
+1960V | 225266.4V | 245.062125 | 100 mA
|
+1970V | 225276.8V | 245.064185 | 100 mA
|
+1980V | 225286.6V | 245.066 | 100 mA
|
+1990V | 225296.1V | 245.06775 | 100 mA
|
+2000V | 225305.1V | 245.0695 | 100 mA
|
+1950V | 225255.5V | 245.06 | 100 mA
|
+1940V | 225246.0V | 245.058125 | 100 mA
|
+1930V | 225236.0V | 245.05625 | 100 mA
|
+1920V | 225226.0V | 245.0545 | 100 mA
|
+1910V | 225216.7V | 245.052625 | 100 mA
|
+1900V | 225205.8V | 245.050375 | 100 mA
|
+1850V | 225156.5V | 245.041 | 100 mA
|
+1800V | 225106.6V | 245.031375 | 100 mA
|
+1950V | 225256.1V | 245.060125 | 100 mA
|
+2000V | 225306.0V | 245.06975 | 100 mA
|
+2050V | 225355.8V | 245.079375 | 100 mA
|
+2100V | 225406.0V | 245.0890 | 100 mA
|
+1950V | 225256.2V | 245.060125 | 100 mA
| Change electron current and note down High Vltage and revolution frequency
NTCAP was restarted at the beginning and stopped at the end of the measuring period.
The NTCAP folder is 05-31_17-16-49 (system time 19:11) |