Message ID: 112     Entry time: Wed Jan 18 13:02:12 2023
Author: Rodolfo 
Category: Laser 
Subject: Testing WS6 with Cobra 
Testing the WS6 (S.N. 3950) together with the Cobra.

Cobra operated with dye DCM in Ethanol + DMSO (Concentration 1:1)
- Oscillator concentration 0.3 g/l.
- Main amplifier concentration 0.0375 g/l (1/8 oscillator concentration).
- Laser linewidth: 6 GHz. Expansion factor: 10x. Grating: 2400 g/mm (40/2013)

Seed energy: 504 mJ @ 532 nm.
Cobra output energy: 112 mJ @ 640 nm.

Scan settings: 640.0 nm - 639.8 nm @ 0.005 nm / step.

Wavemeter settings: Pulsed (ext. trigg. 1 = QRay-Output: Lamp Sync). Precision: Fine.
- Wavemeter calibration. Reference source: Neon lamp.
- LTA-Data under ATPPC023 D:\WLM6 LTA data\D:\WLM6 (18.01.2023, 11.23, 640,10656 nm.lta)
Attachment 1: WS6_640-639.8_0.005_18-01-2023_11.23.png  748 kB  | Hide | Hide all