Dielectronic recombination-assisted laser spectroscopy |
Not logged in |
|
|
Message ID: 112
Entry time: Wed Jan 18 13:02:12 2023
|
Author: |
Rodolfo |
Category: |
Laser |
Subject: |
Testing WS6 with Cobra |
|
|
Testing the WS6 (S.N. 3950) together with the Cobra.
Cobra operated with dye DCM in Ethanol + DMSO (Concentration 1:1)
- Oscillator concentration 0.3 g/l.
- Main amplifier concentration 0.0375 g/l (1/8 oscillator concentration).
- Laser linewidth: 6 GHz. Expansion factor: 10x. Grating: 2400 g/mm (40/2013)
Seed energy: 504 mJ @ 532 nm.
Cobra output energy: 112 mJ @ 640 nm.
Scan settings: 640.0 nm - 639.8 nm @ 0.005 nm / step.
Wavemeter settings: Pulsed (ext. trigg. 1 = QRay-Output: Lamp Sync). Precision: Fine.
- Wavemeter calibration. Reference source: Neon lamp.
- LTA-Data under ATPPC023 D:\WLM6 LTA data\D:\WLM6 (18.01.2023, 11.23, 640,10656 nm.lta) |
|
|