Procedure:
- Scanning phase ends (on LabView wait until "Idle Status" blinks. it will show e.g. 8 loops and 44 scans)
- on DAQ2 (Carsten) and DAQ1 type (simultaneously):
sto ac
- on DAQ2 (Carsten) and DAQ1 type:
clo fi
exit
- on DAQ1 computer, stop GO4 and close it
- on DAQ2 (Carsten) and DAQ1 type:
resl
mbs
@startup
- on DAQ1 computer, start GO4
go4&
- request new injection (move the green slider to red, wait for machine to go further, then move it back to green)
ESR_EXP22_03_209Bi83_Laser_06May.C1.9_WAIT_FOR_MANIP_BY_OP
- in Lassie Monitor, remember the value of SIS-Trafo (GS09DT_ML) after injection, and put it later in the elog table
- wait until the ion beam is cooled (look at the Schottky monitor)
- on DAQ2 (Carsten) and DAQ1 type (simultaneously):
sta ac
- on LabView press on button
Start Scan
- Scanning phase starts.
- you can show the status (and also the current filename) by using the command
sh ac
- update elog with SIS Current, Filenames, and time
- restart from top |