Message ID: 62     Entry time: Thu May 12 15:21:23 2022
Author: Sebastian, Simon 
Category: Laser 
Subject: Flash lamps changed 
We (Sebastian and I) changed the flash lamps of the Quanta Ray Laser and measured the output energy of the laser system afterwards (still during warm up).
All values in [mJ] and over 500 laser-pulses average

DateOut of Quanta Ray @ 532 nmSeed energy for Cobra @ 532 nmOut of Cobra @ 601 nm
10.05.202253348571
11.05.202248448067
12.05.202257850175


The lambda/2 is at 103°. Correction factor for the 532nm-monitor is 11.4 (exchanged broken monitor).