Entry time: Thu May 12 15:21:23 2022
|
Author: |
Sebastian, Simon |
Category: |
Laser |
Subject: |
Flash lamps changed |
|
|
We (Sebastian and I) changed the flash lamps of the Quanta Ray Laser and measured the output energy of the laser system afterwards (still during warm up).
All values in [mJ] and over 500 laser-pulses average
Date | Out of Quanta Ray @ 532 nm | Seed energy for Cobra @ 532 nm | Out of Cobra @ 601 nm
|
10.05.2022 | 533 | 485 | 71
|
11.05.2022 | 484 | 480 | 67
|
12.05.2022 | 578 | 501 | 75
|
The lambda/2 is at 103°. Correction factor for the 532nm-monitor is 11.4 (exchanged broken monitor). |