HOME ESR EXPERIMENTS LABS etc
NucAR lab laser_lab Fritz-Bosch-Lab X-Ray Lab target Beavertail
  Beavertail  Not logged in ELOG logo
Message ID: 15     Entry time: Thu Jun 1 09:39:24 2023
Author: Guy Leckenby, Chris Griffin 
Category: Electronics 
Subject: New new flange test 
We tested the new new flange using the same setup as in ELOG entry 14. The noise level seemed to be somewhere between the middle flange and the old flange.
See attachments:
1. Best noise conditions (grounding two planes and connection to ESR) on worst n-side pad.
2. Worse noise conditions (no grounding) on worst n-side.
3. Best noise conditions on a better n-side. We were unsure why some n-sides were different though.

Thus, for a final comparison of the flanges (best conditions on n-sides):
 - 2016 flange - noise scale = ~5V
 - 2022 flange - noise scale = ~0.4V
 - 2023 flange - noise scale = ~0.8V
Attachment 1: 31.05.23-ESR-NNF-YGC-n-sides.PNG  31 kB  Uploaded Thu Jun 1 10:48:16 2023  | Hide | Hide all
31.05.23-ESR-NNF-YGC-n-sides.PNG
Attachment 2: 31.05.23-ESR-NNF-NGC-n-sides.PNG  33 kB  Uploaded Thu Jun 1 10:48:26 2023  | Hide | Hide all
31.05.23-ESR-NNF-NGC-n-sides.PNG
Attachment 3: 31.05.23-ESR-NNF-YGC-n-sides-v2.PNG  29 kB  Uploaded Thu Jun 1 10:48:46 2023  | Hide | Hide all
31.05.23-ESR-NNF-YGC-n-sides-v2.PNG
ELOG V3.1.5-fc6679b