Entry  Sat Apr 5 14:11:55 2025, CG, GL, ID, General, Run stopped - detectors toast? 
run0034 stopped at 14:11

Leakage currents are increasing and once again we are not fully depleting the pad detectors at 150V.
Entry  Sat Apr 5 15:19:11 2025, Team PLEIADES, General, Run started 
15:14 Started run #35: inside pocket at 20 mm.
Trigger mask changed. Gated only in ch8 in MSI-8 #2 (one side of photodiode)
No events hitting the detector yet.
Entry  Sat Apr 5 17:49:06 2025, CG, GL, ID, Detectors, Outside detector moved in  
After some Sergey-magic the outside pocket moves again. We slowly moved the outside detector in, we are clipping the beam at -20mm. Moved the detector to
-40mm and take some data as of 17:48 (run0036.lmd)
Entry  Mon Apr 7 20:00:13 2025, Guy, General, Results of scope test Screenshot_20250407_172836.pngPicoscope_screenshots.zip
At 10:50am today, we connected a remote Picoscope to the first 4 n-sides, one end of both resistive chains, and both crystal channels to directly measure
the output of the preamps without the ADC clipping at 2V.
Entry  Tue Apr 8 09:53:45 2025, GL, ID, CG, MC, , Moving in Outside Detector 
At 9:53 we moved the outside detector in to take a short data run. We went -110 mm compared to a fully retracted position -126 mm. Data is collected in
run 38 starting at 
9:56:25. Data acquisition was stopped and then moved the detector back at 10:16.
Entry  Tue Apr 8 11:37:30 2025, GL, ID, CG, MC, JG, , Moving in Outside Detector 
We start recording data in run 39 and the outside detector is moved to -100 mm at 11:36. At 11:59 the detector was moved back to the fully retracted position,
-126 mm, and the 
data acquisition was stopped.
Entry  Wed Apr 9 10:38:05 2025, CG, General, End of experiment 
Final injection stored in ring around 10:00 and will be stored for ~1h.
Both detectors are in the parked position but I started a run just as "background"/to see if we get anything.
We have previously seen some events while in this position so who know what might show up.
Entry  Wed Apr 9 11:02:35 2025, Team PLEIADES, Detectors, Beam stopped, voltages off 
11:01 All voltages turned off. Remote control disabled.
Entry  Thu Apr 10 14:58:02 2025, CG, MC, General, Post-run bias curves 
BGO detector assembly originally on flange A.
GAGG  "                        "    flange B.
Entry  Fri Apr 11 14:07:39 2025, CG, General, End of run summary PLEIADES_detector_health.xlsx
All detector components are now in safe storage in Jan's container.

One SiPad (MSPAD 16A) shows lifting of the chip from the PCB in one corner. This has been boxed, labelled and left visible in the lab and should be taken