Message ID: 2     Entry time: Fri Nov 6 09:37:15 2026
Author: Katharina 
Category: Measurement 
Subject: First LaBr calibration 
Energy calibration of LaBr detector with Co and Na spectra measured 28/10/2026 (before this elog was started).
Files: run65.root (22Na), run66.root (60Co).

Fitted gaussian with linear underground to single gamma peaks, pile-up peaks and annihilation peak (only Na).

Calibration plot, see attachment 1.

E(k) = 0.449(1)k - 54.421(5102)
Attachment 1: LaBr_calibration.PNG  101 kB  | Show | Show all