Sat May 17 21:06:52 2025 |
Rodolfo, Christopher, Kristian | Laser | New dye | We have exchanged the dye laser and filled the dye pumps with fresh dye. We also noticed that the seeder was off, we turned on again and set the pump energy to 350 mJ (at 532 nm).
The dye output energy is 150 mJ (at 592 nm) |
85
|
Sat May 17 21:27:17 2025 |
Wilfried, Rodolfo | Laser | Scraping the laser beam | The beam stabilization in not working properly. So we have turned it off. We also scraped the laser beam at the E-Cooler and bring the laser beam to the right position.
The Scraper GECEDS2VU cannot move vertically. Moved scraper horizontally to the inner position. It does not disturb the injection. |
86
|
Sat May 17 21:40:40 2025 |
Wilfried, Rodolfo, Hendrik, Ragandeep, Yuri | Laser | Eveningshift data runs |
Time | File | Ohmlabs | PTB 32 | PTB 33 | ESR intensity SC8 / 1e6 | Typical SIS injections / 1e9 | NTCAP file | comments |
2025-05-17 21:42 | 123 | 205820 | 20582X | 20582X | 6.5 | 3 | SC_2025-05-17_21-41-07 | mbs started, Go4 not connected, only written to disk, wrong scan region, we added 0.27 nm to values but should have subtracted it, see below ** |
| 124 | | | | | | | file empty |
2025-05-17 22:30 | 125 | 20582X | 20582X | 20582X | 6.5 | 2.7 | SC_2025-05-17_21-41-07 | mbs and laser scan started in SC8, before scraping, first and maybe second wavelength affected |
2025-05-17 23:10 | 126 | 20582X | 20582X | 20582X | 0 | 2.7 | SC_2025-05-17_21-41-07 | beam killed by detectors |
2025-05-17 23:42 | 127 | 20582X | 20582X | 20582X | 5 | 2.75 | SC_2025-05-17_21-41-07 |
|
** scan range was incorrect: For file #123 it was set to 593.42 nm - 592.82 nm
*** for #125, set to 592.85 nm - 592.25 nm |
87
|
Sun May 18 00:25:08 2025 |
Finn, Konstantin, Dimitrios | Laser | Nightshift data runs | Took over from afternoon shift and keep scanning
Time | File | Ohmlabs | PTB 32 | PTB 33 | ESR intensity SC8 / 1e6 | Typical SIS injections / 1e9 | NTCAP file | comments |
2025-05-17 23:50 | 128 | 205820 | 20582X | 20582X | 2.6 | 2.8 | SC_2025-05-17_21-41-07 | |
2025-05-17 00:27 | --- | 205820 | 20582X | 20582X | --- | 2.5 | SC_2025-05-17_21-41-07 | Injection was stopped due to radiation alarm at HTD |
2025-05-17 02:01 | 129 | 205820 | 20582X | 20582X | 3.0 | 2.2 | SC_2025-05-17_21-41-07 | beam back (intensity of SIS injection varies) |
2025-05-17 02:36 | 130 | 205827 | 205828 | 205820 | 2.4 | 2.0 | SC_2025-05-17_21-41-07 | |
2025-05-17 03:10 | 131 | --- | --- | --- | --- | --- | SC_2025-05-17_21-41-07 | empty file due to medusa crash |
2025-05-17 03:10 | 132 | 205827 | 20582X | 20582X | 2.4 | 2.2 | SC_2025-05-17_21-41-07 | Started measurement to late, last step affected |
2025-05-17 03:47 | 133 | 205827 | 20582X | 20582X | 3.0 | 2.3 | SC_2025-05-17_21-41-07 | |
2025-05-17 04:24 | 134 | 205827 | 20582X | 20582X | 2.5 | 2.4 | SC_2025-05-17_21-41-07 | |
2025-05-17 05:00 | 135 | 205827 | 20582X | 20582X | 3.9 | 2.4 | SC_2025-05-17_21-41-07 | |
2025-05-17 05:35 | 136 | 205827 | 20582X | 20582X | 1.5 | 2.0 | SC_2025-05-17_21-41-07 | |
2025-05-17 06:10 | 137 | 205827 | 20582X | 20582X | 3.6 | 2.4 | SC_2025-05-17_21-41-07 | |
2025-05-17 06:46 | 138 | 205827 | 20582X | 20582X | 3.2 | 2.9 | SC_2025-05-17_21-41-07 | |
2025-05-17 07:22 | 139 | 205827 | 20582X | 20582X | 3.6 | 1.8 | SC_2025-05-17_21-41-07 | Closed NTCAP file, Nightshift over and out |
|
|
88
|
Sun May 18 01:47:56 2025 |
Konstantin | DAQ | Particle detector | The attachment shows the signal of the particle detector for bunched beam (red, files 104-112) coasting beam (black, files 113-117) as a function of the time for one cycle of the DRALS pattern. One can identify modulations of the signal during the cooling step in both traces. However, there are some clear differences:
- Oscillation in bunched beam operation (maybe due to RF and cooler voltage offset?)
- While in coasting beam mode a steady state is reached, this is not observed in bunched beam operation.
|
89
|
Sun May 18 02:24:46 2025 |
Wilfried, Rodolfo, Carsten | Detectors | Antrieb Cu-Mirror defekt | On Saturday afternoon we moved the Cu mirror to the inner position, however an issue appeared when we tried to move the mirror to the "park" position. The stepping motor did not react properly, it clamped. After restarting the software we could actually move the mirror out, but the stepping motor as well as driving system made a lot of noise until the parking position was reached. This issue has to be investigated specialists after the beamtime. |
90
|
Sun May 18 04:33:46 2025 |
Dimitrios, Finn, Konstantin | Laser | Laser position | We changed the position of the laser back to the position which was set up in E 85 by changing the motor position of motor 4 in the kangoo software at ATPPC031 from -0.1 to -0.3. |
91
|
Sun May 18 08:06:55 2025 |
Danyal, Christopher, Jonas | Runs | morning shift | Took over from night shift and keep scanning
Time | File | Ohmlabs | PTB 32 | PTB 33 | ESR intensity SC8 / 1e6 | Typical SIS injections / 1e9 | NTCAP file | comments |
2025-05-18 08:21 | 140 | 205821 | 205827 | 205828 | 2.4 | 1.6 | SC_2025-05-18_07-57-22 | morning shift starts |
2025-05-18 08:57 | 141 | 205821 | 205828 | 205829 | 3.3 | 1.4 | SC_2025-05-18_07-57-22 | |
2025-05-18 09:32 | 142 | 205821 | 205828 | 205829 | 2.5 | 1.7 | SC_2025-05-18_07-57-22 | |
2025-05-18 10:08 | 143 | 205821 | 205828 | 205829 | 1.3 | 2.0 | SC_2025-05-18_07-57-22 | strong drop in ESR intensity after injection |
2025-05-18 10:43 | 144 | 205822 | 205828 | 205829 | 3.4 | 1.8 | SC_2025-05-18_07-57-22 | |
2025-05-18 11:19 | 145 | 205820 | 205828 | 205829 | 1.1 | 1.5 | SC_2025-05-18_07-57-22 | strong drop in ESR intensity after injection (see screenshot), laser scan started a few seconds too late |
2025-05-18 11:55 | 146 | 205822 | 205828 | 205829 | 0.9 | 1.5 | SC_2025-05-18_07-57-22 | strong drop in ESR intensity after injection, worse than before, decreasing SIS intensity |
2025-05-18 12:00 | | | | | | | | called HKR, reported ESR drop, they checked but see no clear reason, ESR crew will be called for help, see 2nd screenshot |
2025-05-18 12:15 | | | | | | | | Sergey called, we should stop requesting after ESR cycle, then send "ganzen Kontext" (ParamModi) and then request beam again |
2025-05-18 12:29 | 147 | 205822 | 205828 | 205829 | 0.4 | 1.0 | SC_2025-05-18_07-57-22 | again drop in ESR intensity after injection, SIS18 intensity was increased during accumulation |
2025-05-18 12:40 | | | | | | | | we sent "ganzen Kontext" (ParamModi) and then request beam again |
2025-05-18 13:04 | 148 | 205821 | 205828 | 205829 | 1.3 | 2.5 | SC_2025-05-18_07-57-22 | looks better now |
2025-05-18 13:15 | | | | | | | | stopped ESR cycle, stopped NTCAP, Sergey will check ESR settings, maybe improvement is possible |
2025-05-18 13:45 | | | | | | | | ESR cycle running but was interrupted, DRALS was still running, we waited until DRALS end, then started ESR cycle again. We changed time between SIS18 injections from 40 s to 20 s |
2025-05-18 14:14 | 149 | 205826 | 205833 | 205834 | 5.7 | 3.5 | SC_2025-05-18_13-52-38 | the losses after injection (in SC5) are gone!! (see 3rd screenshot) HV values slightly drifted, ~5 V |
2025-05-18 14:49 | 150 | 205819 | 205826 | 205827 | 2.6 | 2.5 | SC_2025-05-18_13-52-38 | slight SC5 losses again, and HV values again slightly drifted, ~3 V |
2025-05-18 15:24 | 151 | 205821 | 205828 | 205829 | 1.8 | 2.3 | SC_2025-05-18_13-52-38 | slight SC5 losses again, HV values back to normal |
2025-05-18 16:00 | 152 | 205824 | 205831 | 205832 | 0.7 | 2.6 | SC_2025-05-18_13-52-38 | low ESR intensity, HV slightly different |
|
|
92
|
Sun May 18 16:08:58 2025 |
Wilfried, Carsten, Rodolfo, Hendrik | Runs | Afternoon shift, continued measurements | Took over from morning shift and keep scanning
Time | File | Ohmlabs | PTB 32 | PTB 33 | ESR intensity SC8 / 1e6 | Typical SIS injections / 1e9 | NTCAP file | comments |
2025-05-18 16:35 | 153 | 205821 | 205827 | 205817 | 0.8 | 2.3 | SC_2025-05-18_13-52-38 | Inconsistent SIS Injections, Voltages all drifted slightly, checked the beam positions by moving the scraper on the remaining ions after scan, Scraper GECEDS2HU was used and the horizontal alignment was confirmed to still be good, Carsten re-sent some SIS commands but changed nothing |
2025-05-18 17:38 | 154 | 205822 | 205828 | 205829 | 0.7 | 2.0 | SC_2025-05-18_13-52-38 | Unlucky SIS injections, Voltages do not drift much anymore, beam after scan was used to scrape again, checked vertical alignment with GECEDS1VU, ion beam located at -0.5, laser beam still clearly visible, around 4mm too high compared to ion beam, GECEDS1HU confirms horizontal alignment is also good like in Scraper 2 |
2025-05-18 18:43 | 155 | 205824 | 205830 | 205831 | 0.9 | 2.5 | SC_2025-05-18_13-52-38 | GECEDS1HU destroyed ion beam at -13.0 |
2025-05-18 19:32 | 156 | 205822 | 205829 | 205830 | 0.8 | 3.0 | SC_2025-05-18_13-52-38 | First few SIS injections were very bad, after that very good, 0.5 on GECEDS2VU and the ion beam vanishes, fits good with laser beam, delay of the laser reduced by 7ns |
2025-05-18 20:21 | 157 | 205822 | 205828 | 205830 | 0.8 | 2.4 | SC_2025-05-18_13-52-38 | |
2025-05-18 20:57 | 158 | 205822 | 205828 | 205830 | 0.7 | 2.5 | SC_2025-05-18_13-52-38 | |
2025-05-18 21:38 | 159 | 205820 | 205826 | 205828 | 1.3 | 2.8 | SC_2025-05-18_13-52-38 | |
2025-05-18 22:13 | 160 | 205820 | 205826 | 205827 | 1.0 | 2.8 | SC_2025-05-18_13-52-38 | |
2025-05-18 22:46 | 161 | 205820 | 205826 | 205827 | 1.7 | 2.9 | SC_2025-05-18_13-52-38 | |
2025-05-18 23:22 | 162 | 205822 | 205829 | 205830 | 3.5 | 2.7 | SC_2025-05-18_13-52-38 | While scanning, at step 16 something weird happened and the accumulated counts decreased only in that step |
2025-05-18 00:00 | 163 | 205822 | 205828 | 205829 | 5.0 | 2.8 | SC_2025-05-18_13-52-38 | 22 injections and 75 breeding cycles |
|
Timing: We shifted the laser approximately 8 ns to an earlier time to compensate for the 2-3 m distance between the target and the detection region (please note that earlier times are located to the right since the x-axis is the time after the event until the common stop). The positions of the laser pulse on the photodiodes are therefore shifted to the right. |
93
|
Sun May 18 21:30:20 2025 |
Wilfried, Rodolfo, Markus, Zoran | Accelerator | T-Scrapers, Update | Here are the right instructions to drive the T-Scrapers inside the ESR Electron Cooler
- ECEDS1VU: Verriegelt mit ECEDS1HA. Der Antrieb kann nur fahren
wenn der Antrieb ECEDS1HA in Position -4,0 steht.
- ECEDS1HA: Verriegelt mit ECEDS1VU. Der Antrieb kann nur fahren
wenn der Antrieb ECEDS1VU in Stellung EI steht.
- ECEDS2VU: Verriegelt mit ECEDS2HA. Der Antrieb kann nur fahren
wenn der Antrieb ECEDS2HA in Position 0,0 steht.
- ECEDS2HA: Verriegelt mit ECEDS2VU. Der Antrieb kann nur fahren
wenn der Antrieb ECEDS2VU in Stellung EI steht.
According to Boywitt/Schuhmacher, BEA-MEI. Date: 28.07.2021 |
94
|
Mon May 19 00:28:40 2025 |
Simon, Finn | Runs | Night shift measurement | Took over from afternoon shift and keep scanning
Time | File | Ohmlabs | PTB 32 | PTB 33 | ESR intensity SC8 / 1e6 | Typical SIS injections / 1e9 | NTCAP file | comments |
2025-05-19 00:20 | 164 | 205822 | 205828 | 205829 | 3.8 | 1.5 | SC_2025-05-18_13-52-38 | week injections from SIS |
2025-05-19 00:50 | 165 | 205822 | 205828 | 205829 | 5.0 | 2.6 | SC_2025-05-18_13-52-38 | |
2025-05-19 01:23 | 166 | 205822 | 205828 | 205829 | 5.3 | 5.3 | SC_2025-05-18_13-52-38 | file empty: medusa crashed |
2025-05-19 01:23 | 167 | ? | ? | ? | --- | --- | SC_2025-05-18_13-52-38 | problem with normalization in Go4: cleaned histograms |
2025-05-19 02:00 | 168 | ? | ? | ? | 0.5 | 2.6 | SC_2025-05-18_13-52-38 | bad injections/stacking |
2025-05-19 02:37 | 171 | ? | ? | ? | 4.5 | 2.8 | SC_2025-05-18_13-52-38 | mbs crashed due to overfull buffer, restarted mbs |
2025-05-19 03:36 | 173 | ? | ? | ? | 5.5 | 2.9 | SC_2025-05-18_13-52-38 | |
2025-05-19 04:14 | 174 | ? | ? | ? | 5.4 | 2.9 | SC_2025-05-18_13-52-38 | |
2025-05-19 04:50 | 175 | ? | ? | ? | 5.3 | 2.8 | SC_2025-05-18_13-52-38 | |
2025-05-19 05:26 | 176 | ? | ? | ? | 5.4 | 2.9 | SC_2025-05-18_13-52-38 | |
2025-05-19 06:02 | 177 | ? | ? | ? | 0.9 | 3.1 | SC_2025-05-18_13-52-38 | great losses after stacking and after breeding |
2025-05-19 06:39 | 178 | ? | ? | ? | 5.1 | 2.9 | SC_2025-05-18_13-52-38 | we changed the scan range to 592.7 - 592.1nm |
2025-05-19 07:15 | 179 | ? | ? | ? | 4.9 | 2.9 | SC_2025-05-18_13-52-38 | |
2025-05-19 07:51 | 180 | ? | ? | ? | 4.9 | 2.9 | SC_2025-05-18_13-52-38 | medusa crashed again |
2025-05-19 08:27 | 181 | ? | ? | ? | 4.9 | 2.8 | SC_2025-05-18_13-52-38 | |
|
After file 168 we noticed that the connection between mbs and medusa looked faulty, so we restarted mbs and restored the connection. But we could not get the voltage divider readout back running.
Files 166,168-172 are empty/incomplete.
After file 177 we saw a peak at the predicted wavelength in the histogram with subtracted background. To confirm this, we changed the scan range to 592.7 - 592.1nm, which should move the peak to the left of the histogram. |
95
|
Tue May 27 10:56:44 2025 |
Konstantin | DAQ | Common stop | The common stop was wired through a 1:2, and subsequently through a 1:99 rate divider channel. However, since the pulse width was too long, the 1:2 channel was translated into a 1:4 reduction. This must be considered, when the data is processed via go4. The reduction factor in the SetParameters.C file must be set to 198 accordingly. |
|