Message ID: 26     Entry time: Thu Jun 6 19:42:20 2024
Author: Carsten 
Category: Detectors 
Subject: Exchange of amplifer channels / cables channels / HV voltage cables for PMT south 
In order to check if the problem of regular burst of pulses on the PMT south is related to amplfier channels and or the HV cable, the cheannels of the amplifiers in the ESR were changed. Likewise, a different HV cable is now used for the supply of the PS.

The problem still persists. It is visible on all three cable lines from ESR to Messhuette (100x DAQ1 cables , and 10x / 100x cables DAQ2).