HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028
  Nuclear hyperfine mixing in 229Th89+ 2024  Not logged in ELOG logo
Message ID: 27     Entry time: Thu Jun 6 19:47:01 2024
Author: Carsten 
Category: DAQ 
Subject: Exchange of Mesytec CFD/shaper - noise on signal lines from ESR magnets 
Jan exchanged the Mesytec CFD/shper since it could not be "remote controlled" anymore. The new one has different shaping times and delays.
It is noted the threshold range is rather small. 
With the new  shaper only the x100 signals could meaningfully discriminated etc.
After start of the ESR (magnets etc.) the signals of the 10x channels see quite some noise that makes it more difficult to adjust the thresholds for the low signals.
ELOG V3.1.5-fc6679b