Category: Laser 
Subject: Wavemeter HeNe long time stability 
The HeNe was running over night and the measured laser frequency of the wavemeter was logged, see picture.
A significant drift of around 300 MHz is visible during the night.
It is possible, that the laser lab got too cold during the night, as the pulsed laser was not running.
For now, we put the wavemeter in a thermoinsulator box...