HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028
  Nuclear hyperfine mixing in 229Th89+ 2024  Not logged in ELOG logo
Message ID: 69     Entry time: Thu Jun 13 18:23:46 2024
Author: Julian 
Category: Laser 
Subject: Restarting the laser controller 
If the laser controller is locked up again, the little display will show an error message after an attempted medusa scan.
To restart it, follow these steps:
1.) Unlock the beam stabilization.
2.) shut down medusa and the wavemeter software
3.) Turn off the 2 laser stages
4.) Close the shutter of the pumping laser
5.) Turn off and on the power button of the pulsed laser inside of the black box of the pump laser insertion.

To start again go back through the list.
ELOG V3.1.5-fc6679b