Category: Laser 
Subject: Restarting the laser controller 
If the laser controller is locked up again, the little display will show an error message after an attempted medusa scan.
To restart it, follow these steps:
1.) Unlock the beam stabilization.
2.) shut down medusa and the wavemeter software
3.) Turn off the 2 laser stages
4.) Close the shutter of the pumping laser
5.) Turn off and on the power button of the pulsed laser inside of the black box of the pump laser insertion.

To start again go back through the list.