Message ID: 77     Entry time: Fri Jun 14 14:44:51 2024
Author: Afternoon shift 
Category: Runs 
Subject: Afternoon shift 
Pizza shift

the scan direction is from higher wavelength to lower wavelength, even though it is noted otherwise here in the software.
From hardware side, the laser can only scan in one direction: long wave length to short wave length, independent of what is written in the tables.


Time Part. Det. Int ESR-intensity cooler-HV LMD-DAQ1 Files LMD-DAQ2 Files scan range(nm) number of positions dwell time(s) n loops comments
14.06.2024 12:45 38900 2e5 104411 72 89 549.6 - 550.4 201 1s 3 starting with 50 accumulation
14.06.2024 13:13 41400 2e5 104411 73 90 549.6 - 550.4 201 1s 3 there was a :10 crash during the injection, laser scan started normally.
14.06.2024 13:41 39500 2e5 104411 74 91 549.6 - 550.4 201 1s 3
14.06.2024 14:38 67750 2e5 104412 75 92 549.6 - 550.4 201 1s 3 afternoon shift takes over. After short break from hkr to optimize beam, proceeding with 30 injections, particle number detector might not have been reset, check next run
14.06.2024 15:02 34500 2e5 104412 76 93 549.6 - 550.4 201 1s 3
14.06.2024 15:38 27500 1e5 104412 77 94 549.6 - 550.4 201 1s 3
14.06.2024 16:01 25300 1e5 104412 78 95 549.6 - 550.4 201 1s 3
14.06.2024 16:23 28700 1e5 104412 79 96 549.6 - 550.4 201 1s 3 At around 550.2 nm the wavemeter is overexposed
14.06.2024 16:48 xxx xx xx 80 97 xx xx xx xx empty file , no laser scan, work on laser, while trying to adjust the wavemeter, we tried to move the laser using medusa, at some point, this caused the cobra to go into error mode again, onlz a restart could solve it. As a consequence, we still do not know if the wavemeter coupling is optimal
14.06.2024 17:17 30000 1e5 104412 81 98 549.6 - 550.4 201 1s 3 wavemeter coupling looks good now
14.06.2024 17:41 56900 1e5 104412 82 99 549.6 - 550.4 201 1s 3 ratemeter was probably not reset
14.06.2024 18:04 8100 1e5 104412 83 100 549.6 - 550.4 201 1s 3 really bad injection (looked fine (?)), low ion count from SIS, they exchanged a cathode @UNILAC
14.06.2024 18:42 3200 3e4 104412 84 101&102(automatically opened new file without command after 1GB?) 549.6 - 550.4 201 1s 3 problems with accelerator: after deacceleration the beam is gone. Scraper position has moved(?). The cooling water from GSI is too warm, so we had to shut down the laser. NTCAP restarted. And NTCAP restarted again @20:39 for test data for Shahan.