Nuclear hyperfine mixing in 229Th89+ 2024 |
Not logged in |
|
|
Message ID: 9
Entry time: Sat Jun 1 18:00:20 2024
|
Author: |
Carsten |
Category: |
DAQ |
Subject: |
ReadOut Trigger im Timer Scaler changed because of faulty signal from frequency generator |
|
|
Old readout trigger (~200 Hz) in 1MHz Scaler showed two nearby peaks indicating a jitter of the readout frequency or same faulty output driver of the frequency generator Metrix GX240 used to produce the readout frequency (see below).
Used the 1 kHz clok instead divided by 5 -> only a single peak in 1MHz spectrum in the scaler (see below).
This should be used during the beamtime. |