HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  Nuclear hyperfine mixing in 229Th89+ 2024  Not logged in ELOG logo
Message ID: 9     Entry time: Sat Jun 1 18:00:20 2024
Author: Carsten 
Category: DAQ 
Subject: ReadOut Trigger im Timer Scaler changed because of faulty signal from frequency generator 
Old readout trigger (~200 Hz) in 1MHz Scaler showed two nearby peaks indicating a jitter of the readout frequency or same faulty output driver of the frequency generator Metrix GX240 used to produce the readout frequency (see below).


Used the 1 kHz clok instead divided by 5 -> only a single peak in 1MHz spectrum in the scaler (see below).



This should be used during the beamtime.
ELOG V3.1.5-fc6679b