Message ID: 9     Entry time: Sat Jun 1 18:00:20 2024
Author: Carsten 
Category: DAQ 
Subject: ReadOut Trigger im Timer Scaler changed because of faulty signal from frequency generator 
Old readout trigger (~200 Hz) in 1MHz Scaler showed two nearby peaks indicating a jitter of the readout frequency or same faulty output driver of the frequency generator Metrix GX240 used to produce the readout frequency (see below).


Used the 1 kHz clok instead divided by 5 -> only a single peak in 1MHz spectrum in the scaler (see below).



This should be used during the beamtime.