HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038 E0075 E0203_E0018 E0052_2025
  Nuclear hyperfine mixing in 229Th89+ 2025  Not logged in ELOG logo
Message ID: 23     Entry time: Wed Jun 18 08:07:17 2025
Author: Julien, Konstantin 
Category: Laser 
Subject: new FCU Calibration 
UV power was not stable but dropping from ~18 mW to 13 mW after opening COBRA shutters.
New calibration of frequency conversion unit:

Wavelength/nm       Position/Steps
554,0001            383029
553,0000            386604
552,0001            390229
551,0000            393954
550,0001            397679
549,0001            401404
548,0001            405104
547,0001            408829

now stable at ~19-20 mJ depending on wavelength

Table of dataset: Attachement 1
Plot of dataset: Attachement 2
Attachment 1: 2025_06_18_FCU_Cal_Table.JPG  23 kB  Uploaded Wed Jun 18 09:17:40 2025  | Hide | Hide all
2025_06_18_FCU_Cal_Table.JPG
Attachment 2: 2025_06_18_FCU_Cal_Plot.JPG  44 kB  Uploaded Wed Jun 18 09:18:00 2025  | Hide | Hide all
2025_06_18_FCU_Cal_Plot.JPG
ELOG V3.1.5-fc6679b