  Nuclear hyperfine mixing in 229Th89+ 2025  Not logged in ELOG logo
Message ID: 53     Entry time: Mon Jun 23 10:59:36 2025
Author: Julian, Lukas, Zsolt, Nadine, Konstantin 
Category: Runs 
Subject: Scan routine 
-Start new injection
->wait for 60 injections to be complete
->note down SIS intensity (at extraction)
->observe Schottky to ensure proper bunching
->check laser position (should be well-centered)
-open file, start Go4
->do '@openfile' in both mbs windows
->start acquisition (green arrow) in both Go4 instances (check indicator 'Current Ev/s' at the bottom, should be green, not red)
->start laser scan
->reset Ch3 of the particle detector (in the electronics hut)
->check laser position (disappears at start of scan then drifts through the image from the left)
-when laser scan is finished:
->stop Go4 acquisition (big red 'STOP' button in the top row)
->do '@closefile' in both mbs windows
->check counts on particle detector
-update elog
-repeat