  laser spectroscopy on Be-like Kr  Not logged in ELOG logo
Message ID: 44     Entry time: Fri Feb 28 00:19:04 2025
Author: Konstantin, Sebastian 
Category: Detectors 
Subject: PMT power supply 
We checked the PMT signal inside the cave by using the local HV power supply. If one needs to repeat the measurement at another time, be careful to use the correct HV power supply, since the second one, which is mounted in the crate has the wrong polarity. Both PMT's must be supplied with a negative voltage.
During our tests the detector signals looked quite well and the background rate was below 10Hz. The baseline oscillated by 2mV (20mV after first amplification).

Subsequently, we did the tests of Entry 37 again. CH00 of the HV supply in the ESR container shows a higher leakage current of around 485muA at 2650V compared to 156muA at Ch01. Both channels could hold the voltage and the rate was measured with the LED at the XUV-detector operated a volage of 6V. See attachment.