HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038 E0075
  laser spectroscopy on Be-like Kr  Not logged in ELOG logo
Message ID: 57     Entry time: Fri Feb 28 20:44:04 2025
Author: Konstantin 
Category: DAQ 
Subject: Updated Channel list 
Final (hopefully) channel list:

TDC Channels
Channel # Kanalbelegung
0 Laser busy Kr
1 Next Laserstep Kr
2
3
4 PMT Mitte
5 PMT Nord
6 XUV Anode
7
8 Shutter open
9 Shutter close
10 Bunch HF (common stop)
11 Shutter auf
12 Shutter zu
13 PD NW / 89
14 PD SW / 89
15 AFG rep / 89



Scaler Channels
Channel # Kanalbelegung
16 DC Trafo
17 Injection
18 U-ecool
19 I-ecool
20 P-gun
21 P-coll
22 common stop (Bunch HF)
23 1MHz clock
24 Extraktion
25 Shutter (10kHz)
26
27
28
29
30
31


Shutter open signal attached at channel 8 is used as reference for the measurement window. This can be adapted in line 28 of the SetParameters.C file in the folder '/u/lascool/2025laserspectroscopy/kr_analysis/'.
Example:

fPar->piSoftwaregateLength = 20000; // How many CS after Shutter open used for signal in 4_LaserSteps/signal.

Currently, the common stop has a rate of 1.7838817 MHz divided by 89. This corresponds to approximately 50mus per common stop. With the setting of the example, the measurement window would be 1s after the shutter has opened.