Entry  Mon Feb 17 13:49:20 2025, Denise, General, Frequenzbreiten Ionensystem 
Berechnung der Frequenzbreiten des Lasers im Ionensystem 
Werte für die Frequezbreiten des Lasers aus Dissertation Benedikt Langfeld übernommen

v=0.644c 

Pulsdauer = 115ps
Frequenzbreite (FWHM) = 10.87GHz
Frequenzbreite Ionensystem = 23.36GHz

Pulsdauer = 240ps
Frequenzbreite (FWHM) = 4.19GHz
Frequenzbreite Ionensystem = 9.00GHz

Pulsdauer = 400ps
Frequenzbreite (FWHM) = 2.82GHz
Frequenzbreite Ionensystem = 6.06GHz

Pulsdauer = 735ps
Frequenzbreite (FWHM) = 2.05GHz
Frequenzbreite Ionensystem = 4.41GHz
Entry  Fri Feb 14 16:41:54 2025, Sebastian , General, Verkabelung PD ESR IMG_8368.jpeg
Verkabelung von den Photodioden im ESR. Sind auch gelabelt.

PD gehen zuerst einmal in den FAMP (10x Verstärkung) 

PD NW -> ESR Panel 4 
PD SW -> ESR Panel 7 

XUV LED -> ESR Panel 8 
Entry  Fri Feb 14 16:24:22 2025, Denise, General, Strahlprofil UV im ESR  IMG_8360.jpeg796a74f2-ec76-4396-82e5-9d3209b36612.jpeg0bbd839a-f6cf-4fa4-a5a4-71d0c84044fa.mov
Erstes Bild ist das Strahlptofil in Höhe des Nord 
towers. 
Zweites Bild (mit Millimeterpapier) und Video zeigt 
das Strahlprofils etwa in Höhe des XUV Detektors im 
ESR. Das Zittern ist dabei durch anstoßen am Nord 
Tower entstanden, wichtig hier zu sehen ist das 
„Atmen“ des Strahlprofils, welches dabei allerdings 
örtlich fest bleibt. 
Entry  Thu Feb 13 12:56:01 2025, Danyal, General, ESR access - monday (whole day) 8-18 hr 
 
Entry  Thu Feb 13 12:55:13 2025, Danyal, General, ESR access - tomorrow (Fri) 10:00 - 12:00 
 
Entry  Wed Feb 12 12:48:48 2025, Danyal, General, ESR access  
We had the chance to shortly enter the ESR, between 10:45 and 11:30. 

- We repositioned the PSDs of the TEM system. 
- We could see the laser beam at the ESR exit window. 
- We used a wedge at the exit window and directed the two laser beams onto a power meter and a fast photodiode. 
- We photographed the XUV detector setup. 
- We documented the cabling at the central cable panel. 

In the ESR laser lab we could see the laser beam power on the power meter and we could also see 
the laser pulses on the fast photodiode (scope). 

We still need a longer break (~2 hrs) to further optimize the setup and all the settings. 
Entry  Tue Feb 11 13:08:32 2025, Danyal, Accelerator, limited access to ESR this week 
Hitrap will kein ESR Zugang erlauben bevor 22 Uhr. Außerdem merken die, wenn ESR-SIS nicht laufen, dass sie morgens immer 1 Stunde Zeit verlieren weil nicht "angefordert" wurde. Deswegen wollen sie, dass ESR-SIS nachts auch weiter Strahl anfordern, etwa 1 mal pro 5 Minuten. Aber wir finden schon noch eine Lücke um reinzugehen.


Hitrap does not want to allow ESR access before 10 p.m. In addition, when ESR-SIS is not running, they notice that they always lose 1 hour of time in the morning because nothing was "requested". That's why they want ESR-SIS to continue requesting beams at night, about once every 5 minutes. But we can still find a gap to go in.
Entry  Tue Feb 11 10:35:38 2025, Danyal, Detectors, XUV detector 
Volker Hannen kommt am 19.2 (Mittags) und dann machen wir Abends den XUV Detektor bereid.


Volker Hannen is coming on February 19th (in the afternoon) and then we'll get the XUV detector ready in the evening.
Entry  Tue Feb 11 10:32:01 2025, Danyal, General, access to ESR 
Es wird noch zwei 2 stündige Unterbrechungen am UNILAC geben. Wahrscheinlich diesen Freitag und nächste Woche Mittwoch oder Donnerstag. Am 25.2. wird dann die Gleitringdichtung der Pumpe am UNILAC gewechselt (8-18 Uhr) daher rechne ich mit Xe Strahl im ESR frühestes am 26.2.
Von HITRAP habe ich erfahren, dass wir Abends/Nachts was im Ring machen dürfen, ab 22 Uhr. z.B. am 17.2, 19.2 und 21.2.  Aber ich denke vorher (diese Woche) geht es bestimmt auch mal.

There will be two 2-hour breaks at the UNILAC. Probably this Friday and next week Wednesday or Thursday. On February 25th the mechanical seal of the pump on the UNILAC will be changed (8 a.m. to 6 p.m.) so I expect the Xe beam in the ESR on February 26th at the earliest.
I found out from HITRAP that we can do something in the ring in the evening/night, from 10 p.m. onwards. For example on February 17th, February 19th and February 21st.  But I think there will be some possibilities already this week. 
Entry  Tue Feb 11 10:30:21 2025, Danyal, General, laser valves are open 
Die Laserventile im ESR sind nun offen. Der Laserstrahl darf durchgeschossen werden.

The "laser valves" (NW and SW) are now open.  The laser beam may pass through the ESR. 
Entry  Thu Feb 6 12:54:29 2025, Danyal, Accelerator, HV-divider for electron cooler connected - readout is running HV_divider_readout.jpg
The HV-divider has been connected to the E-Cooler. 

The readout (by Konstantin) is running too. 

Konstantin will prepare the inclusion into our DAQ.
Entry  Thu Feb 6 11:03:26 2025, Danyal, General, pressures in laser window sections NW and SW 
The pressures, read by the UHV technician, are : 

E01VVLF1T (E01VMLF1X):           5.84e-10mbar
E02VVLF2T (E02VMLF2X):           2.91e-10mbar

In 2019 the pressures were sth like 2e-9 mbar. They are much lower now. 

Note: The valves LF1 and LF2 may only be opened after ESR extraction to HITRAP has been set up! 
Most likely this is Monday 10 Feb. 


The UHV technician is: 
Mr. Semjon Strohmenger 
Phone: +49 6159 71 1765 
PSA: 7651 
Entry  Wed Feb 5 14:05:53 2025, Rodolfo, Sebastian, Analysis, Schirm / Iris ESR Entrance Fenster 
Wegen defekt von Kippschirm wurde das alte System zum Verstellen der Iris und des Schirms für das ESR Eintrittsfensters genutzt.

Fleischmann Trafo:

Slot 1 -> Schirm verfahren
Slot 3 -> Iris verfahren
Entry  Wed Feb 5 13:50:28 2025, Sebastian, Accelerator, TIF Modul - Injektions-/ Extraktionssignale 
TIF Moduleinstellung
Remark: Die EV H und EV L Einstellungen sind die Eventnummern in Hexadezimal Basis 16

Injektionssignal
Super Zyklus          3 (steht für ESR)
MA                    2 (Maschine 2)
EV H                  3
EV L                  1 


Extraktionssignal (Richtung Hitrap) 
Super Zyklus          3 (steht für ESR)
MA                    7 (steht für Subchain 7)
EV H                  4 (steht für Event 72 in hexadezimal = 48)
EV L                  8 (steht für Event 72 in hexadezimal = 48)

Das Extraktionssignal muss für unseres ESR Pattern noch angepasst werden.
Der "Code" 3748 wurde nur genutzt, um für die Laser - DAQ Einstellungen das Extraktionssignal zu erhalten.
Prinzip: Nach Extraktion wird der Laser auf einen Wellenlängenschritt weiter eingestellt.
Entry  Tue Feb 4 13:01:52 2025, Sebastian , Accelerator, Druck NW und SW Laserfenster IMG_8500.jpegIMG_8499.jpeg
 
Entry  Fri Jan 31 16:37:39 2025, Sebastian, DAQ, PD Nord IMG_8478.jpeg
PD Nord an ESR Panel PMT Süd verkabelt
Entry  Wed Jan 29 10:23:21 2025, Sebastian, General, Remote Control Schirm Nord West Tower 
Box zum rein und rausfahren hinter Laser PC im ESR

Zeiger von Rädchen muss auf "SCHIRM NW" stehen

Knopf oben betätigt den Schirm zum rein und rausfahren.


Zusätzlich gibt es im ESR am NW Tower einen Knopf zwischen Magnet und Tower, um den Schirm dort auch rein und rausfahren zu können.
Entry  Tue Jan 28 15:57:05 2025, Denise, General, Justage: ESR Labor Periskop Spiegel oben 
obere Schraube nach rechts drehen, dann bewegt sich Laserspot nach links 
selbe Schraube nach links dann Laser rechts

untere Schraube nach rechts drehen dann geht Laser nach unten 
selbe Schraube Laser nach oben 
Entry  Tue Jan 28 14:16:22 2025, Sebastian, General, Shutter 
Thorlabs Shutter SH1/M kann über STRPC001 (ESR Laserlab PC) und der Software SC10 genutzt werden.

COM 9 für Connection
Baud Rate 9600
Entry  Tue Jan 28 14:13:54 2025, Benedikt & Sebastian TEAMWORK DIGGA, DAQ, RedPitaya Adresse 
SCPI Server:
http://169.254.204.79/

Port wenn nötig: 5000

Wurde auf SISNBG002 gestartet, da direkt am Tower STRPC001 nicht möglich war den RedPitaya unter 
rp-f0b7a4.local/ zu erreichen