Message ID: 210
Entry time: Sat Apr 5 18:24:09 2025
|
Author: |
Dima, Zac, Shahab |
Category: |
DAQ |
Subject: |
RSAs stop again, due to misses |
|
|
RSAs had to stop because the cycle of 1.8 came too often, both analyzers could not catch up with the file transfer.
changed to 1.7 seconds
last file in the 1.8 setting:
RSA01-2025.04.05.18.20.46.005
RSA02-2025.04.05.18.20.43.346
new filenames from the 1.7 s settings:
RSA01-2025.04.05.18.28.12.146.tiq
new analysis settings:
[analysis]
nframes = 80
lframes = 131072
avg = 4
Explanation:
every file has 10625050 samples with 1.7 seconds length
sampling rate: 6.25 MSPS
this means every sample is worth: 1.6e-7 seconds
this means every frame with length of 2^17 samples will have a time length of: 21 ms
finally with an averaging of 4 frames into one, this means that every time frame in the NPZ file will have 80ms length
frequency setting
please see attached pictures. |
|
|
|
|