  281   Mon Apr 7 17:07:33 2025 Team PLEIADESDetectorsMoving OUTSIDE PLEIADES detector slowly in (and out)
17:08 Tried to move outside particle detector (GE01DD2AG) in - it is stuck again :-(
17:18 Sergey tried to move it, now it moves again. Moved to -100 mm.
17:30 Moved detector to outside parking position
  306   Tue Apr 8 09:53:04 2025 Team PLEIADESDetectorsMoving outside particle detector in for 20min
9:53 Data taking with outside detector (GE01DD2AG) moved in to -110mm
  308   Tue Apr 8 10:13:51 2025 Team PLEIADESDetectorsMoved outside particle detector back into parking position
10:16 Moved outside particle detector back into parking position (-126mm)
  310   Tue Apr 8 11:37:13 2025 Team PLEIADESDetectorsPLEIADES
11:37 Outside particle detector (GE01DD2AG) moved in to -100mm to take data
Attachment 1: 2025-04-08_11-38-37-375.png
2025-04-08_11-38-37-375.png
  311   Tue Apr 8 11:57:56 2025 Team PLEIADESDetectorsOutside detector moved out
12:00 Outside particle detector moved out to parking position again (-125 mm)
  9   Mon Mar 31 16:53:03 2025 ShahabDAQstop and delete NTCAP data from Weekend test
stop and delete NTCAP data from Weekend test

for now the NAS data2 bank has been flushed to make sure good operation for today's beam.
we have now 35 TB space on bank 2.

the recording will be done on the bank2 of the NAS.

There are still some junk data on bank1, which we will remove later.
  10   Mon Mar 31 17:46:21 2025 EstherDAQNTCAP scaler assignment
The setup contains two NI-6612 I/O cards with BNC-2121 connector accessories that provide eight inputs each for TTL scaler signals.
Important: When connecting scaler signals make sure that they are TTL (not NIM)!

Additionally, there is the possibility to assign data streams (eg clocks) from other modules in the NI chassis to scaler channels.
Below you find the assignment of the channels on the software side, including the internal clocks. This is important when analysing the .tmds files.

ChannelInputComment
010 MHz clockexternal (from NI 6674T timing module)
1DC trafoion current in ESR
2injection triggerfrom Genesys (IO2 on sddsc221)
3U-coolercooler voltage
4I-coolercooler electron current
5P-gunpressure at e-cooler
6100 kHz clockinternal
720 MHz clockinternal
8software triggerfrom Detectomer
9PLEIADES outside
9PLEIADES inside
  12   Mon Mar 31 20:53:02 2025 Shahab, Carlo, Zac, Esther, MichaelDAQDAQ start
20:28 started NTCAP
started queqiao file transfer for RSAs only
started analyzer looper for RSAs

 
  21   Tue Apr 1 02:55:06 2025 Shahab, Michael, Yue, Zac, CarloDAQnew RSA recording
anything from RSA01-2025.04.01.02.54.25.976.tiq onwards is good data

recording parameters:

5 MHz with 6.25 MSPS, 2.5 seconds, analyzer span 100 khz, trigger at 5%

All analyzer times are now set to summer time
file time stamps are correct

settings are in the attached pictures.
Attachment 1: IMG_4955.JPG
IMG_4955.JPG
Attachment 2: IMG_4956.JPG
IMG_4956.JPG
Attachment 3: IMG_4954.JPG
IMG_4954.JPG
  23   Tue Apr 1 03:50:12 2025 Shahab, Yue, Carlo, Zac, Esther, MichaelDAQTuners of the both cavities set
410 MHz cavity resonance frequency tuned. motors both on 50 mm
245 MHz cavity resonance frequency tuned. motors both on 390 mm
  24   Tue Apr 1 03:52:56 2025 Shahab, Michael, Yue, Zac, CarloDAQgood files
all settings are done.
any files coming after this one is good:

RSA01-2025.04.01.03.52.59.247.tiq
  42   Tue Apr 1 15:32:59 2025 ZacDAQNTCAP Restarted
NTCAP Has been restarted (we restart every day around 15:00 so data goes in a new folder)
  50   Tue Apr 1 20:37:04 2025 JeroenDAQChange of settings for shorter acq. window
Acquisition stopped at 20:36. 

First file from this setting: RSA01-2025.04.01.03.52.59.247.tiq

Last file from this setting: RSA01-2025.04.01.20.36.54.257.tiq

acq. length changed to 500 ms

acq bw changed to 50 MHz

==> 12 MSamples

Trigger window condition set to 50%.

First file from this setting: RSA01-2025.04.01.20.45.40.053.tiq

Analysis setting the the frame length 2^16 and frame number = 190.
  51   Tue Apr 1 21:05:20 2025 Jeroen, Zac, Esther, Shahab, WolframDAQChange of settings for shorter acq. window
> Acquisition stopped at 20:36. 
> 
> First file from this setting: RSA01-2025.04.01.03.52.59.247.tiq
> 
> Last file from this setting: RSA01-2025.04.01.20.36.54.257.tiq
> 
> acq. length changed to 500 ms
> 
> acq bw changed to 50 MHz
> 
> ==> 12 MSamples
> 
> Trigger window condition set to 50%.
> 
> First file from this setting: RSA01-2025.04.01.20.45.40.053.tiq
> 
> Analysis setting the the frame length 2^16 and frame number = 190.

Last file from previous setting: RSA01-2025.04.01.21.06.30.580.tiq

Changed the RSA01,02 to 50 MSamples/s (20nsec/sample) for 500.00 ms acq. length, 25 MHz acq. BW, 33% trigger position.

First good file from this time: 21:37 04.01.2025 -- RSA01-2025.04.01.21.37.13.721.tiq
  52   Tue Apr 1 22:03:07 2025 Jeroen, Zac, Esther, Shahab, WolframDAQOptimisation of settings
Estimation of potential gain by changing. Files from this points are going to rubbish. Potential gains are c. 20% 

22:04: The current acquisition window is recording for 2.9 s. And is not recording for 3.9 s. We are running currently with a spill cycle of 5s every repetition.

22:22 The spill cycle between each injection was changed to 4->3->2s progressively and FRS also stopped taking spills so we settled on 2s spill request.

In the morning the spill request has to be changed back to 5s when the FRS goes back into operation.

new file with the setting is

RSA01-2025.04.01.22.19.06.761.tiq

[analysis]
nframes = 190
lframes = 65536
navg = 1
  55   Tue Apr 1 23:07:15 2025 ShahabDAQshort interruption of queqiao script for RSAs
has been restored.
  57   Tue Apr 1 23:56:18 2025 Shahab, Carlo, Zac, Yue, MichaelDAQChange of RSA01 settings for new harmonic=212.
23:55 stop the RSAs. 

410:
last file from previous setting (harmonic=211): 2025.04.01.23.55.21.712.
new plunger position for 410 cavity: 33.5 inside and outside.
new center frequency: 410535000 Hz.

245:
harmonic didn't change, still 126.
adjusted center frequency to 243997000 Hz.

00:31 restart the RSAs.
new files from: 2025.04.02.00.31.11.796.
  61   Wed Apr 2 02:51:31 2025 Carlo, Zac, ChrisDAQBeam is back
02:44 beam is back
  62   Wed Apr 2 04:33:57 2025 Carlo, Zac, ChrisDAQScreenshots Lassie
screenshots of Lassie
Attachment 1: 2025-04-02_04-28-59-944_tcl1064.png
2025-04-02_04-28-59-944_tcl1064.png
Attachment 2: 2025-04-02_03-38-54-877_tcl1064.png
2025-04-02_03-38-54-877_tcl1064.png
  65   Wed Apr 2 07:16:33 2025 Carlo, Chris, Alex, MichaelDAQESR not requesting
7:05 ESR not requesting new pulses from the SIS. 
Operators are trying to fix the problem.

7:20 eam is back