Entry  Sat Apr 5 17:57:08 2025, Ragan, Dima, Zac, Shahab, Calibration, New Settings Resonators 
New settings for resonators:

410: plunger stay the same
center: 411.756955 MHz


245: plunger now at 150
center: 245.507668 MHz


first files on the resonators:


RSA01-2025.04.05.18.09.08.519.tiq
RSA02-2025.04.05.18.09.09.618.tiq
Entry  Sat Apr 5 17:45:17 2025, Ragan, Shahab, Dima, Zac, DAQ, restarted NTCAP 
NTCAP restarted 17:43

we have 63% full on bank 2 of the NAS
which means about 11% every day
this gives us roughly another 3 days of recording time
we can always switch to bank 1, but changing that, needs changing the analysis paths. so for now, we continue on bank2

:-)
Entry  Sat Apr 5 17:34:14 2025, Team PLEIADES, Detectors, Moving outside detector in  
17:35 Slowly moving PLEIADES outside detector (GE01DD2AG) in for some data. Clipping ions at -20mm. Leaving detector at -40mm for a few minutes.
Entry  Sat Apr 5 17:26:48 2025, Team PLEIADES, Detectors, Outside pocket "repair" 
We tried to convince the outside pocket to move, no success. We can hear the "hissing sound" of the pressurized air but no movement. Sergey tried to move it from the HKR, also no success. After that I tried to move it again from the Messhuette...and the "Out of tolerance" error message is gone, and the pocket seems to move. (We cannot confirm from inside because the ESR is already closed).
Guess the outside pocket got intimidated by Sergey and decided to get moving again... ;-)

We will take a bit more beam once the setting is stable, but try not to fry another detector.
Entry  Sat Apr 5 17:01:30 2025, Ragan, Dima, Zac, Shahab, DAQ, DAQs stopped 
NTCAP and analysers are stopped at 17:00.Iris goes inside to move the detectors 
Entry  Sat Apr 5 16:54:58 2025, Zsolt, General, 188Hf +72 setting back 
The 188Hf72+ setting measurement starts now. Exactly the same setting like two days ago.
Entry  Sat Apr 5 16:48:26 2025, Ragan, Wolfram, Sergey, Zsolt, Yuri, Rui-Jiu, General, 188Hf72+ 
Now ESR is set to the 188Hf72+ setting.
Entry  Sat Apr 5 16:24:07 2025, Shahab, Dima, Zac, DAQ, restarted looper script for RSA02 
we removed the wrong entries from the PKL file, and restarted the looper script for RSA02.

important change:

nframes = 84 
lframes = 131072
navg = 4



here are the new parameters:

# Config file for the multiprocessing looper script

[paths]
monitor_dir = "/data.local/G22-00018_00203/data/analyzers/RSA02"
output_dir = "/data.local/G22-00018_00203/results/analyzers/RSA02"
state_file = "/data.local/G22-00018_00203/scripts/e0018_e0203_scripts/processed_files_rsa2.pkl"

[processing]
num_cores = 12
interval_seconds = 0.5
file_ready_seconds = 0.5

[analysis]
nframes = 84 
lframes = 131072
navg = 4
zzmin = 0
zzmax = 1e6
mask = false
dbm = false
todo = ['spectrum', 'spectrogram'] # choose any of or combination 'png', 'spectrum', 'spectrogram'
Entry  Sat Apr 5 15:45:48 2025, Ruijiu, Ragan, Analysis, See single ion of 186Hf72+ and isomer in ntcap data Screenshot_2025-04-05_15-45-26.pngScreenshot_2025-04-05_15-55-38.pngScreenshot_2025-04-05_16-03-51.png
The ntcap data record 410 detector continuously. The frame length is  26 ms/frame * 32 frames= 832 ms.
Entry  Sat Apr 5 15:25:16 2025, Shahab, DAQ, stop RSA02 looper script 
2025-04-05 15:25
Stop the RSA02 looper script to change the settings of analysis
Entry  Sat Apr 5 15:20:19 2025, Yuri, General, Change to 188Hf  
Before changing we check how much we cut with the ESR scrapers
two lines
208Pb80+ with slits intensity 15 arb u
204Hg78+ ... 0.055

Now, slits fully open


208Pb80+ with slits intensity 127 arb u
204Hg78+ ... 0.25

Increase compared to 188Hf measurement:
208Pb80+  increase of factor 8.5
204Hg78+  increase of factor 14
Entry  Sat Apr 5 15:15:33 2025, Xinliang Yan, Wenwen Ge, General, Lassie screen shots 2025-04-05_15-14-09-343.png2025-04-05_15-13-46-936.png2025-04-05_15-15-04-554.png2025-04-05_15-14-44-712.png
 
Entry  Sat Apr 5 15:01:03 2025, Hannes, Cagla, General, Lassie screenshots 2025-04-05_15-01-39-374.png2025-04-05_15-01-22-352.png2025-04-05_15-01-59-451.png2025-04-05_15-02-06-386.png
 
Entry  Sat Apr 5 14:55:21 2025, Team PLEIADES, Detectors, Moving detectors in again, to 20mm 
14:57 We are moving the inside pocket detector in again, this time to 20mm
Entry  Sat Apr 5 14:11:49 2025, Team PLEIADES, Detectors, Moving inside pocket detector 
14:12 We are moving the inside pocket out for a moment, change the bias
Entry  Sat Apr 5 14:02:54 2025, Hannes, Cagla, General, Lassie screenshots 2025-04-05_14-03-09-807.png2025-04-05_14-02-50-834.png2025-04-05_14-03-58-456.png2025-04-05_14-03-52-527.png
 
Entry  Sat Apr 5 12:59:08 2025, Hannes, Cagla, General, Lassie screenshots 2025-04-05_12-59-26-722.png2025-04-05_12-59-07-111.png2025-04-05_13-00-24-333.png2025-04-05_13-00-15-781.png
 
Entry  Sat Apr 5 12:22:12 2025, Team PLEIADES, Detectors, PLEIADES inside detector in 
12:23 moved PLEIADES inside detector in to 2mm.
Entry  Sat Apr 5 12:03:23 2025, Hannes, Cagla, General, Lassie screenshots 2025-04-05_12-02-57-892.png2025-04-05_12-02-51-503.png2025-04-05_12-03-22-319.png2025-04-05_12-03-41-597.png
 
Entry  Sat Apr 5 11:37:44 2025, Ragan, Rui-Jiu, Zsolt, Wolfram, Yuri, General, General 
We have not seen 184Hf and 186Hf so far. 