ID |
Date |
Author |
Category |
Subject |
57
|
Tue Apr 1 23:56:18 2025 |
Shahab, Carlo, Zac, Yue, Michael | DAQ | Change of RSA01 settings for new harmonic=212. |
23:55 stop the RSAs.
410:
last file from previous setting (harmonic=211): 2025.04.01.23.55.21.712.
new plunger position for 410 cavity: 33.5 inside and outside.
new center frequency: 410535000 Hz.
245:
harmonic didn't change, still 126.
adjusted center frequency to 243997000 Hz.
00:31 restart the RSAs.
new files from: 2025.04.02.00.31.11.796. |
12
|
Mon Mar 31 20:53:02 2025 |
Shahab, Carlo, Zac, Esther, Michael | DAQ | DAQ start |
20:28 started NTCAP
started queqiao file transfer for RSAs only
started analyzer looper for RSAs
|
9
|
Mon Mar 31 16:53:03 2025 |
Shahab | DAQ | stop and delete NTCAP data from Weekend test |
stop and delete NTCAP data from Weekend test
for now the NAS data2 bank has been flushed to make sure good operation for today's beam.
we have now 35 TB space on bank 2.
the recording will be done on the bank2 of the NAS.
There are still some junk data on bank1, which we will remove later. |
55
|
Tue Apr 1 23:07:15 2025 |
Shahab | DAQ | short interruption of queqiao script for RSAs |
has been restored. |
93
|
Wed Apr 2 20:57:34 2025 |
Shahab | Detectors | 10GB cables for the DAQ rack |
We have been working together with IT on the speed of data transfer: with the new 10 GB interface we can now transfer data from NAS to Lustre with 400MB/s. From NAS to GPU Computer up to 700MB/s. All cables around the rack are now 10GB capable. |
137
|
Fri Apr 4 01:12:03 2025 |
Shahab | Analysis | NPZ add up script |
Since there are many NPZ summer scripts around, I thought I would also share mine. This one takes advantage of the power of filelists and has a progress bar. Creating file lists gives you more control over the range of analyzed data, for example, the command:
find /data.local/G22-00018_00203/results/analyzers/RSA02 -name "RSA02-2025.04.04.01.??.??.???.tiq_spectrogram.npz" > filelist.txt
makes a file list of everything file that was produced during the second hour of the morning of the April 4th. You can fine tune the wildcards according to the time range you are interested in.
then just call the script by:
python npz_addup.py filelist.txt
you can also provide time cuts, like this:
python npz_addup.py --time-cut 0.14 filelist.txt
which will cut away first 0.14 seconds of the spectrogram. Without time cuts, it will do over the whole range.
It creates 4 type of files:
summed_spectrum.npz --> for use with RionID for PID
summed_spectrogram.npz --> for general use / analysis
summed_spectrum.png --> for general view
summed_spectrogram.png --> for detailed view
if time-cut is provided, the filenames will have a "_time_cut" suffix at the end.
Here is the link to the script:
https://github.com/xaratustrah/e0018_e0203_scripts/blob/main/npz_addup.py
examples are in attachment: |
200
|
Sat Apr 5 15:25:16 2025 |
Shahab | DAQ | stop RSA02 looper script |
2025-04-05 15:25
Stop the RSA02 looper script to change the settings of analysis |
214
|
Sat Apr 5 20:58:37 2025 |
Shahab | DAQ | removed temporary NPZ files for additional disk space |
Deleted 14 TB of unused spectrum.npz files from APPC, so that there is more disk space.
switched off the production of .tiq_spectrum.npz files to save disk space in the looper script
Explanation:
.tiq_spectrogram.npz —> 3D NPZ files, that can be used for add up and create small 2D "spectrum" file for RionID
.tiq_spectrum.npz —> 2D NPZ files with super high resolution, noone needs them up to now |
277
|
Mon Apr 7 16:07:42 2025 |
Shahab | General | all config files |
here are all TOML config files of the automatic scripts. |
285
|
Mon Apr 7 17:51:15 2025 |
Shahab | General | HowTo change resonance frequency of 410 MHz Cavity |
step by step pictures:
sometimes the name of the motor connection is not available in the menu, it should be added manually.
Add MBOX:
put some name, like "410 MHz Schottky Cavity in ESR"
put the IP address: 140.181.139.194
press connect
change to local control
press initialize
go to motor drive
go to motor 1 and 2
always move same value for both motors. |
290
|
Mon Apr 7 18:58:04 2025 |
Shahab | DAQ | changed averaging in NPZ files |
new analysis settings for NPZ files:
[analysis]
nframes = 190
lframes = 65536
navg = 1
first file with this setting:
'RSA02-2025.04.07.18.57.17.454.tiq', 'RSA02-2025.04.07.18.57.02.328.tiq', 'RSA02-2025.04.07.18.58.17.958.tiq', 'RSA02-2025.04.07.18.57.40.143.tiq', 'RSA02-2025.04.07.18.58.25.520.tiq',
'RSA02-2025.04.07.18.57.09.892.tiq', 'RSA02-2025.04.07.18.57.32.580.tiq', 'RSA02-2025.04.07.18.57.25.017.tiq', 'RSA02-2025.04.07.18.58.10.395.tiq', 'RSA02-2025.04.07.18.57.55.271.tiq',
'RSA02-2025.04.07.18.56.54.765.tiq', 'RSA02-2025.04.07.18.58.02.833.tiq', 'RSA02-2025.04.07.18.57.47.707.tiq', 'RSA02-2025.04.07.18.56.47.202.tiq' |
330
|
Fri Apr 11 15:49:41 2025 |
Shahab | General | Screenshots of RSAs |
Screenshots of RSAs showing acquisition settings, for documentation, they show the time / date on the bottom left, in case needed for data analysis. |
331
|
Fri Apr 11 15:57:37 2025 |
Shahab | General | More screenshots of RSAs |
more screen shots of RSAs. |
322
|
Wed Apr 9 00:08:30 2025 |
Sergey, Yuri and Xinliang | General | Move of the scraper positions at ESR |
From roughly … to 0:05AM, Sergey moved GEEXDS1HA (not
changed
in the end) and GE01DS3HG (changed from -85 to -80 mm) to reduce
the acceptance of the ESR to improve the mass resolution.
After the change, there are less small traces appearing in the
413.55735 MHz +/-5 kHz range.
The final positions of the two scrapers are
GEEXDS1HA :-10 mm; GE01DS3HG:-80 mm. |
5
|
Fri Mar 28 22:20:47 2025 |
Sergey, Shahab | General | Pattern |
the ACC pattern for both experiments. |
321
|
Tue Apr 8 23:08:15 2025 |
Sergey Litvinov, Xinliang Yan | General | Tuning of the ion optics |
From roughly 10:45 pm to 11:05 pm, Sergey tuned the
ion optics of the ESR to see if the mass resolution at
413.55735 MHz improves or not. Please note that the
frequencies of ions during the tuning process are not
stable.
Q-magnets and S-magnets have been adjusted, see
accelerator E-logs ID-844 titled ‘ ESR seting for
204Pt77+’ created by Sergey. |
73
|
Wed Apr 2 10:02:40 2025 |
Sergey + Yuri | Accelerator | ESR Timeout |
At around 9:40, ESR timeout changed from 10 s to 30 s
Triggering only on real injections.
Before that, every second injection was not real but the old one. They can be identified since they should have no "kicker" disturbance. |
84
|
Wed Apr 2 16:28:14 2025 |
SS, ID, HA etc | DAQ | NTCAP stop/start |
NTCAP was stopped and restarted. |
121
|
Thu Apr 3 13:36:17 2025 |
Ruijiu,Yuri, Wolfram,Zsolt | Analysis | 188Hf identification |
|
158
|
Fri Apr 4 16:24:48 2025 |
Ruijiu, Yuri | Analysis | PID of 186Hf |
Code used to add up injection data.
(pyroot) gpuuser@appc256:/data.local/G22-00018_00203/scripts/OnlineDataAnalysisSystem/shift/20250331/data/2025.04.04.13.00$ find /data.local/G22-00018_00203/results/analyzers/RSA01 -name "RSA01-2025.04.04.*.tiq_spectrogram.npz" > filelist.txt
(pyroot) gpuuser@appc256:/data.local/G22-00018_00203/scripts/OnlineDataAnalysisSystem/shift/20250331/data/summed_spectrum$ python npz_addup.py --time-cut 0.14 filelist.txt |