HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 258     Entry time: Mon Mar 23 18:15:55 2020
Author: Laszlo 
Category: General 
Subject: noisy 90degree Xray det 
Year:  
Uwe found out, that there was some common ground effect between 90 and 145 degree detectors. after moving the grounds, we could record much more nicer (Gaus 
shaped) peaks in the spectra. Now we continue with this settings
ELOG V3.1.5-fc6679b