HOME
ESR
EXPERIMENTS
LABS etc
|
E121
E127
E132
E125
E143
laser_cooling_2021
E142
Ê128
E146
E0052
E0018
E0028
E0038
|
proton-capture on 118Te |
Not logged in |
|
|
Message ID: 270
Entry time: Sun Apr 5 20:34:42 2020
|
Author: |
Laszlo |
Category: |
Calibration |
Subject: |
Xray energies - 1. calibration |
Year: |
|
|
|
30angle: E[keV] = 0.0158507 *ch-2.07456
90angle: E[keV] = 0.01950*ch-2.76
145angle: E[keV] = 0.017988 *ch-1.809
I could identify 2 peaks in the 241Am spectra and 4 peaks in the 133Ba. These peaks are fitted with a gaussian to get their position --> linear energy calibration for the combined (Am+Ba) data sets for each detector. The function I used: E(ch) [keV] = m*ch+b |
|
|
|
|
|
|
|
|
|