HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 270     Entry time: Sun Apr 5 20:34:42 2020
Author: Laszlo 
Category: Calibration 
Subject: Xray energies - 1. calibration 
Year:  
30angle:  E[keV] = 0.0158507 *ch-2.07456
90angle:  E[keV] = 0.01950*ch-2.76
145angle: E[keV] = 0.017988  *ch-1.809

I could identify 2 peaks in the 241Am spectra and 4 peaks in the 133Ba. These peaks are fitted with a gaussian to get their position --> linear energy calibration for the combined (Am+Ba) data sets for each detector. The function I used: E(ch) [keV] = m*ch+b
Attachment 1: 90angle_Xray_first_calibration.png  52 kB  Uploaded Sun Apr 5 21:34:55 2020  | Hide | Hide all | Show all
90angle_Xray_first_calibration.png
Attachment 2: 30angle_Xray_first_calibration.png  53 kB  Uploaded Sun Apr 5 21:34:59 2020  | Show | Hide all | Show all
Attachment 3: 145angle_Xray_first_calibration.png  51 kB  Uploaded Sun Apr 5 21:35:02 2020  | Hide | Hide all | Show all
145angle_Xray_first_calibration.png
Attachment 4: 90_det_calib_Am_Ba.cal  462 Bytes  Uploaded Sun Apr 5 21:35:11 2020  | Show | Hide all | Show all
Attachment 5: 30_det_calib_Am_Ba.cal  522 Bytes  Uploaded Sun Apr 5 21:35:19 2020  | Show | Hide all | Show all
Attachment 6: 145_det_calib_Am_Ba.cal  522 Bytes  Uploaded Sun Apr 5 21:35:26 2020  | Show | Hide all | Show all
Attachment 7: run041_133Ba_high_90.svg  35 kB  Uploaded Sun Apr 5 21:35:40 2020  | Show | Hide all | Show all
Attachment 8: run040_241Am_90.svg  33 kB  Uploaded Sun Apr 5 21:38:12 2020  | Hide | Hide all | Show all
Cannot create thumbnail, please check ImageMagick installation