On 30. and 31.03.2021 we did some testing with the 6 BaF detectors.
Trigger & threshold:
With a threshold at ~250 keV, each detector has an trigger rate of about 0.5kHz due to internal activity.
Energy spectra
All detectors showed the Na22 lines (511 + 1275 keV) and also the lines from internal activity.
The energy resolution was on the order of 8.5 - 11 keV, depending on the detector.
Time spectra
Raw TDC spectra showed self-stop peak.
TDC diff spectra (between detectors) show one main (sharp) correlation peak about 8ns = 30ch*0.293ns/ch off the zero.
At very low intensity a time structure around this peak is visible in a region of 120ns around the main peak. This is not understood, the energy spectra look similar for all these time regions (above/on/below the main peak).
Additionally, the coincidence E-specta between detector 1 and 2 (gating on the main time peak), nicely showed only the 511 keV line from Na22 and some gras.
Overnight test run:
Det. 1, 2, 3, 4 with the Na22 source
lxg1275:/data.local3/e127/lmd_2021/test/
e127b_run0021.lmd
e127b_run0022.lmd
e127b_run0023.lmd
e127b_run0024.lmd
e127b_run0025.lmd
e127b_run0026.lmd
e127b_run0027.lmd
e127b_run0028.lmd
e127b_run0029.lmd
e127b_run0030.lmd
e127b_run0031.lmd
e127b_run0032.lmd
e127b_run0033.lmd
e127b_run0034.lmd
Gain matched HV settings: https://elog.gsi.de/esr/E127/289
Issues to solve:
- 2 missing HV channels
- Go4 Analysis not stable
- trigger not on Baf but on Si |