Message ID: 290
Entry time: Wed Apr 14 09:42:50 2021
|
Author: |
Jan |
Category: |
Detectors |
Subject: |
BaF HV gain matching |
Year: |
2021 |
|
|
The preliminary gain matching settings for a full ADC range of ~12MeV:
Det.Nr. | ADC/TDC ch | HV | I | comment
|
1 | 20 | 2400V | 390uA | CAEN HV ch0
|
2 | 21 | 1763V | 287uA | CAEN HV ch1
|
3 | 22 | 2340V | 380uA | CAEN HV ch2
|
4 | 23 | 2175V | 354uA | CAEN HV ch3
|
5 | 24 | 2017.5V | ? | emetron HV
|
6 | 25 | 1780V | ? | not yet biased
|
Attached is a 22Na-spectrum taken with detectors 1 - 5.
The used MSCF settings are also attached.
Detector 6 does not yet have a HV channel available. |
|
|