HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 290     Entry time: Wed Apr 14 09:42:50 2021
Author: Jan 
Category: Detectors 
Subject: BaF HV gain matching 
Year: 2021 
The preliminary gain matching settings for a full ADC range of ~12MeV:

Det.Nr. ADC/TDC ch HV I comment
1 20 2400V 390uA CAEN HV ch0
2 21 1763V 287uA CAEN HV ch1
3 22 2340V 380uA CAEN HV ch2
4 23 2175V 354uA CAEN HV ch3
5 24 2017.5V ? emetron HV
6 25 1780V ? not yet biased


Attached is a 22Na-spectrum taken with detectors 1 - 5.
The used MSCF settings are also attached.

Detector 6 does not yet have a HV channel available.
Attachment 1: BaF_Na22_gain_match_v1.png  26 kB  Uploaded Wed Apr 14 10:49:53 2021  | Show | Show all
Attachment 2: baf_gain_match_v1.dat  1 kB  Uploaded Wed Apr 14 10:55:38 2021  | Show | Show all