Efficiency calibration in the lab
Detector: GEM1800 - 90 deg
Source: 133Ba_high
Distance: 167.5mm
Start time: 14:34:51 - 30.04.2021
Stop time: 14:45:42 - 30.04.2021
file name: e127b_run0010.lmd
avrg. rate: 2600 Hz
dead-time: 16%
Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt
Threshold at ~10keV
We have some external noise on the signal. |