|   proton-capture on 118Te  | 
Not logged in | 
  | 
   | 
| 
Message ID: 304
    Entry time: Sat May  1 22:34:16 2021
 |  
| Author: | 
Jan Glorius  | 
 | Category: | 
Calibration  | 
 | Subject: | 
efficiency xray 90deg  | 
 | Year: | 
2021  | 
 
  |  
 
 | 
the attached efficiency curve for the 90 deg. xray detector is made with the runs
e127b_run0008.lmd
e127b_run0009.lmd
e127b_run0010.lmd
All of them were taken in the lab (not in ESR).
This is a very preliminary analysis!
E.g. deadtime has been estimated and was not cleanly determined.  | 
 | 
 | 
59.54	0.00212	0.00007
26.34	0.00180	0.00011
53.16	0.00215	0.00007
79.61	0.00218	0.00008
81.00	0.00222	0.00007
34.95	0.00209	0.00007
35.87	0.00209	0.00009
30.80	0.00199	0.00006
46.54	0.00217	0.00009 
 |