HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 307     Entry time: Tue May 4 14:34:51 2021
Author: Jan Glorius 
Category: Detectors 
Subject: Xray weight control induces signals 
Year: 2021 
When the readout/control of the det. weight control via serial connection is active, there is a clear signal induced in the Xray detectors for each command sent.

The solution for now is, that we activate weight control NOT DURING MEASUREMENT!
ELOG V3.1.5-fc6679b