HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 322     Entry time: Wed May 12 10:16:10 2021
Author: Jan Glorius 
Category: Calibration 
Subject: run0026 - Xray145 calib Ba133_hi d=305mm 
Year: 2021 
Efficiency calibration in the lab

Detector: GEM3000 - 145 deg
Source: 133Ba_hi
Distance: 305mm
Start time: 10:15:51 - 12.05.2021
Stop time:  10:59:35 - 12.05.2021

file name: e127b_run0026.lmd
avrg. rate: 850 Hz
dead-time:  5%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
ELOG V3.1.5-fc6679b