Message ID: 330     Entry time: Thu May 13 11:51:42 2021
Author: Jan Glorius 
Category: Calibration 
Subject: run0034 - Xray90 calib Am241_hi d=167.5mm 
Year: 2021 
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************

Detector: GEM1800 - 90 deg
Source: 241Am_hi
Distance: 167.5mm
Start time: 11:49:53 - 13.05.2021
Stop time:  12:02:24 - 13.05.2021

file name: e127b_run0034.lmd
avrg. rate: 600Hz
dead-time:  5%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.