HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 332     Entry time: Thu May 13 12:21:41 2021
Author: Jan Glorius 
Category: Calibration 
Subject: run0036 - Xray90 calib Ba133_lo d=167.5mm 
Year: 2021 
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 90deg distance
But still operated as Xray1 in DAQ
************************************

Detector: GEM1800 - 90 deg
Source: 133Ba_lo
Distance: 167.5mm
Start time: 12:21:21 - 13.05.2021
Stop time:  13:22:04 - 13.05.2021

file name: e127b_run0036.lmd
avrg. rate: 300Hz
dead-time:  2%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.