HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 334     Entry time: Thu May 13 14:16:58 2021
Author: Jan Glorius 
Category: Calibration 
Subject: run0038 - Xray145 calib Pb210 d=305mm 
Year: 2021 
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Rate is increasing slowly, from ~70 to ~220 Hz in ~7 hours.
Also the threshold seems to be lower now.
Also possible source holder movement!
Resolution still good.
Take this run with care!

Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 14:16:18 - 13.05.2021
Stop time:  21:09:12 - 13.05.2021

file name: e127b_run0038.lmd
avrg. rate: 70Hz > 220Hz!!!!
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
ELOG V3.1.5-fc6679b