Efficiency calibration in the lab
************************************
CAUTION!
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************
Rate is increasing slowly 200 > 400 Hz, more noise coming in.
Also possible source holder movement!
Resolution still good.
Take this run with care!
Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 21:22:00 - 13.05.2021
Stop time: 08:56:31 - 14.05.2021
file name: e127b_run0039.lmd
avrg. rate: 220Hz > 400Hz !!!!
dead-time: 3%
Blocking gate 2.5 us in vulom.
Threshold < 10keV
The scale serial communication is off. |