HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 336     Entry time: Thu May 13 21:22:31 2021
Author: Jan Glorius 
Category: Calibration 
Subject: run0039 - Xray145 calib Pb210 d=305mm 
Year: 2021 
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Rate is increasing slowly 200 > 400 Hz, more noise coming in.
Also possible source holder movement!
Resolution still good.
Take this run with care!

Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 21:22:00 - 13.05.2021
Stop time:  08:56:31 - 14.05.2021

file name: e127b_run0039.lmd
avrg. rate: 220Hz > 400Hz !!!!
dead-time:  3%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
ELOG V3.1.5-fc6679b