Message ID: 337     Entry time: Fri May 14 09:06:55 2021
Author: Jan Glorius 
Category: Calibration 
Subject: run0040 - Xray145 calib Ba133_hi d=305mm 
Year: 2021 
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Threshold slightly increased

Detector: GEM1800 - 145 deg
Source: 133Ba_hi
Distance: 305mm
Start time: 09:06:33 - 14.05.2021
Stop time:  09:31:32 - 14.05.2021

file name: e127b_run0040.lmd
avrg. rate: 850Hz
dead-time:  6%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.