HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 338     Entry time: Fri May 14 09:35:13 2021
Author: Jan Glorius 
Category: Calibration 
Subject: run0041 - Xray145 calib Cd109 d=305mm 
Year: 2021 
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

Threshold slightly increased

Detector: GEM1800 - 145 deg
Source: 109Cd
Distance: 305mm
Start time: 09:34:52 - 14.05.2021
Stop time:  09:47:42 - 14.05.2021

file name: e127b_run0041.lmd
avrg. rate: 600Hz
dead-time:  5%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
ELOG V3.1.5-fc6679b