HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 341     Entry time: Fri May 14 13:58:52 2021
Author: Jan Glorius 
Category: Calibration 
Subject: run0044 - Xray145 calib 210Pb d=305mm 
Year: 2021 
Efficiency calibration in the lab

************************************
CAUTION! 
This is now GEM1800 @ 145deg distance
But still operated as Xray1 in DAQ
************************************

90deg TFA diff set from 50 to 200.
Also threshold is slightly increased.

Detector: GEM1800 - 145 deg
Source: 210Pb
Distance: 305mm
Start time: 13:58:33 - 14.05.2021
Stop time:  16:46:00 - 14.05.2021

file name: e127b_run0044.lmd
avrg. rate: 40Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
ELOG V3.1.5-fc6679b