HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te  Not logged in ELOG logo
Message ID: 71     Entry time: Fri Mar 20 04:29:45 2020
Author: Jan 
Category: DAQ 
Subject: DAQ settings for the first night 
Year:  
Attachment 1: e127.trlo  7 kB  | Show | Hide all | Show all
Attachment 2: main.cfg  1017 Bytes  | Hide | Hide all | Show all
#log_level = verbose

CRATE("Master")
{
  event_max = 1
  	BARRIER
  	MESYTEC_MADC32(0x00A10000) {
		  nim_busy = gate0
   	          range = 8 V
		  resolution=8
		  hires=true
	}
	BARRIER
	MESYTEC_MADC32(0x00A20000) {
                  nim_busy = gate0
                  range = 4 V
                  resolution=8
                  hires=true
		  }
	#BARRIER
        #CAEN_V965 (0x00B10000) {
	#	  channel_enable = 0..11
        #}
	BARRIER
	CAEN_V775 (0x00C10000) {
		  common_start = false
		  time_range = 1200 ns
		  suppress_invalid = true
		  suppress_over_range = true
	}
	BARRIER
        CAEN_V775 (0x00C20000) {
                  common_start = false
                  time_range = 1200 ns
                  suppress_invalid = true
                  suppress_over_range = true
        }
	BARRIER
	#CAEN_V830 (0x00E10000) {
	#	  resolution = 32	
	#	  }
	#BARRIER
	#CAEN_V830 (0x00E20000) {
        #          resolution = 32
	#}
	#BARRIER
	GROUP("Timetamp")
	GSI_VULOM(0x05000000) {
		timestamp = true
	}
}
Attachment 3: r3bfuser.cfg  916 Bytes  | Hide | Hide all | Show all
[r3bfuser]
# White Rabbit ID, =0 -> don't write MBS TO timestamp sub-event.
# Subevent 10/1: WR subsystem ID + timestamp
# wr_id = 0x100

# Save LOS/ROLU scalers/samplers every n:th spill, =0 -> don't read.
# Subevent 38/3800: ECL_IN scalers + pileup histogram
# Subevent 39/3900: trloii sampler data
do_beam_scalers = 1

# Save spill structure with 1/10s sized bins.
# do_spill_struct = 10

# Handle spill triggers 10..13 specially, >0 -> keep-alive grace period in s.
# period put to ~2 weeks, to enable long-term vulom scalers with targetOFF
do_spill_triggers = 1000000

# Save TPAT sub-event for every event
# Subevent 36/3600: trlo multi-event TPAT information
do_tpat = 1

# Save LMU scalers on spill triggers (10..13).
# Subevent 37/3700: lmu scalers (input, before / after DT & after reduction)
do_lmu_scalers = 1

# Send beam samplers and LMU scalers via UDP, !="" -> target host.
do_udp = "lxg1275.gsi.de"
Attachment 4: setting.1584673412  5 kB  | Show | Hide all | Show all
ELOG V3.1.5-fc6679b