HOME
ESR
EXPERIMENTS
LABS etc
|
E121
E127
E132
E125
E143
laser_cooling_2021
E142
Ê128
E146
E0052
E0018
E0028
E0038
|
proton-capture on 118Te |
Not logged in |
|
|
Message ID: 71
Entry time: Fri Mar 20 04:29:45 2020
|
Author: |
Jan |
Category: |
DAQ |
Subject: |
DAQ settings for the first night |
Year: |
|
|
|
|
|
|
#log_level = verbose
CRATE("Master")
{
event_max = 1
BARRIER
MESYTEC_MADC32(0x00A10000) {
nim_busy = gate0
range = 8 V
resolution=8
hires=true
}
BARRIER
MESYTEC_MADC32(0x00A20000) {
nim_busy = gate0
range = 4 V
resolution=8
hires=true
}
#BARRIER
#CAEN_V965 (0x00B10000) {
# channel_enable = 0..11
#}
BARRIER
CAEN_V775 (0x00C10000) {
common_start = false
time_range = 1200 ns
suppress_invalid = true
suppress_over_range = true
}
BARRIER
CAEN_V775 (0x00C20000) {
common_start = false
time_range = 1200 ns
suppress_invalid = true
suppress_over_range = true
}
BARRIER
#CAEN_V830 (0x00E10000) {
# resolution = 32
# }
#BARRIER
#CAEN_V830 (0x00E20000) {
# resolution = 32
#}
#BARRIER
GROUP("Timetamp")
GSI_VULOM(0x05000000) {
timestamp = true
}
}
|
|
[r3bfuser]
# White Rabbit ID, =0 -> don't write MBS TO timestamp sub-event.
# Subevent 10/1: WR subsystem ID + timestamp
# wr_id = 0x100
# Save LOS/ROLU scalers/samplers every n:th spill, =0 -> don't read.
# Subevent 38/3800: ECL_IN scalers + pileup histogram
# Subevent 39/3900: trloii sampler data
do_beam_scalers = 1
# Save spill structure with 1/10s sized bins.
# do_spill_struct = 10
# Handle spill triggers 10..13 specially, >0 -> keep-alive grace period in s.
# period put to ~2 weeks, to enable long-term vulom scalers with targetOFF
do_spill_triggers = 1000000
# Save TPAT sub-event for every event
# Subevent 36/3600: trlo multi-event TPAT information
do_tpat = 1
# Save LMU scalers on spill triggers (10..13).
# Subevent 37/3700: lmu scalers (input, before / after DT & after reduction)
do_lmu_scalers = 1
# Send beam samplers and LMU scalers via UDP, !="" -> target host.
do_udp = "lxg1275.gsi.de"
|
|