  proton-capture on 118Te, Page 14 of 29  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Category Subject Yeardown
  547   Sat May 29 18:29:26 2021 Sophia, MarioRunsRun 01982021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0198.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 18:28:12 29.05.2021
run stop: 19:33:30 29.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  548   Sat May 29 19:34:41 2021 Sophia, MarioRunsRun 01992021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0199.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 19:33:33 29.05.2021
run stop: 20:33:33 29.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  549   Sat May 29 20:35:22 2021 Sophia, MarioRunsRun 02002021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0200.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 20:33:36 29.05.2021
run stop: 20:45:12 29.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  550   Sat May 29 21:37:54 2021 Sophia, MarioRunsRun 02012021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0201.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 21:01:17 29.05.2021
run stop: 22:02:02 29.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  551   Sat May 29 22:03:28 2021 Sophia, MarioRunsRun 02022021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0202.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 22:02:07 29.05.2021
run stop: 22:24:21 29.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  552   Sat May 29 23:05:52 2021 Oliver, DiegoGeneralNo Beam2021
Due to a radioprotection issue in CaveA there is no beam. We wait for the radioprotection shift to arrive...

Beam back at 23:18
  553   Sat May 29 23:18:43 2021 Oliver, DiegoRunsRun 02032021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0203.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 23:18:19 29.05.2021
run stop: 00:18:53 30.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 410 nA


SIS particles before: 2.0-2.9e9 (one shoot)
ESR current at injection: 1.8e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  554   Sun May 30 00:14:58 2021 Oliver, DiegoGeneralSIS Pattern changed2021
The pattern in SIS was changed so that we get more injections into ESR. Before HTD had 9 seconds which resulted in a timeout in the ESR. Now HTD has 10 seconds which improved a lot...
  555   Sun May 30 00:19:33 2021 Oliver, DiegoRunsRun 02042021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0204.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 00:18:57 30.05.2021
run stop: 01:21:30 30.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 100 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 400 nA


SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.5e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  556   Sun May 30 01:22:03 2021 Oliver, DiegoRunsRun 02052021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0205.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 01:21:34 30.05.2021
run stop: 02:25:32 30.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 100 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 410 nA


SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.75e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  557   Sun May 30 02:26:07 2021 Oliver, DiegoRunsRun 02062021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0206.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 02:25:37 30.05.2021
run stop: 03:30:03 30.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 130 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 420 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 2.75e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  558   Sun May 30 03:30:37 2021 Oliver, DiegoRunsRun 02072021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0207.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 03:30:06 30.05.2021
run stop: 04:27:27 30.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 120 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 420 nA


SIS particles before: 2.6e9 (one shoot)
ESR current at injection: 2.9e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  559   Sun May 30 04:27:58 2021 Oliver, DiegoRunsRun 02082021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0208.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 04:27:30 30.05.2021
run stop: 05:32:04 30.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 100 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 420 nA


SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.5e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  560   Sun May 30 05:32:42 2021 Oliver, DiegoRunsRun 02092021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0209.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 05:32:09 30.05.2021
run stop: 06:32:13 30.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 110 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 410 nA


SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.5e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  561   Sun May 30 06:32:43 2021 Oliver, Diego, JanRunsRun 02102021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0210.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 06:32:16 30.05.2021
run stop: 07:12:04 30.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 110 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 400 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 2.7e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  562   Sun May 30 07:26:14 2021 Oliver, Diego, JanRunsRun 02112021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0211.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm

run start: 07:25:54 30.05.2021
run stop: 08:04:19 30.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 110 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 400 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 2.7e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13
  563   Sun May 30 10:24:20 2021 Jan GloriusCalibrationrun0212 - X-calib. 90-pb 145-am2021
Efficiency calibration in ESR after experiment

************************************
GEM1800 @ 90deg (xray1) 
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************


Detectors: 90 + 145 
Source @ 90deg: 210Pb
Source @ 145deg: 241Am_hi
Distance: each
Start time: 10:24:55 - 30.05.2021
Stop time:  11:19:29 - 30.05.2021

file name: e127b_run0212.lmd
avrg. rate: 250Hz
dead-time:  2%
  564   Sun May 30 11:32:06 2021 Jan GloriusCalibrationrun0213 - X-calib. 90-am 145-pb 35-ba2021
Efficiency calibration in ESR after experiment

************************************
GEM1800 @ 90deg (xray1) 
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************


Detectors: 90 + 145 + 35 
Source @ 90deg: 241Am_hi
Source @ 145deg: 210Pb
Source @ 35deg: 133Ba_hi
Distance: each
Start time: 11:31:47 - 30.05.2021
Stop time:  13:10:14 - 30.05.2021

file name: e127b_run0213.lmd
avrg. rate: 700Hz
dead-time:  6%
  565   Sun May 30 13:23:26 2021 Jan GloriusCalibrationrun0214 - X-calib. 90-ba 145-ba 35-am2021
Efficiency calibration in ESR after experiment

************************************
GEM1800 @ 90deg (xray1) 
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************


Detectors: 90 + 145 + 35 
Source @ 90deg: 133Ba_lo
Source @ 145deg: 133Ba_hi
Source @ 35deg: 241Am_hi
Distance: each
Start time: 13:22:38 - 30.05.2021
Stop time:  14:28:09 - 30.05.2021

file name: e127b_run0214.lmd
avrg. rate: 1150Hz
dead-time:  8%
  566   Sun May 30 14:38:09 2021 Jan GloriusCalibrationrun0216 - X-calib. 90-ba 35-pb2021
Efficiency calibration in ESR after experiment

************************************
GEM1800 @ 90deg (xray1) 
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************


Detectors: 90 + 35 
Source @ 90deg: 133Ba_lo
Source @ 35deg: 210Pb
Distance: each
Start time: 14:46:46 - 30.05.2021
Stop time:  20:11:08 - 30.05.2021

file name: e127b_run0216.lmd
avrg. rate: 330Hz
dead-time:  3%