ID |
Date |
Author |
Category |
Subject |
Year |
563
|
Sun May 30 10:24:20 2021 |
Jan Glorius | Calibration | run0212 - X-calib. 90-pb 145-am | 2021 |
Efficiency calibration in ESR after experiment
************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************
Detectors: 90 + 145
Source @ 90deg: 210Pb
Source @ 145deg: 241Am_hi
Distance: each
Start time: 10:24:55 - 30.05.2021
Stop time: 11:19:29 - 30.05.2021
file name: e127b_run0212.lmd
avrg. rate: 250Hz
dead-time: 2% |
564
|
Sun May 30 11:32:06 2021 |
Jan Glorius | Calibration | run0213 - X-calib. 90-am 145-pb 35-ba | 2021 |
Efficiency calibration in ESR after experiment
************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************
Detectors: 90 + 145 + 35
Source @ 90deg: 241Am_hi
Source @ 145deg: 210Pb
Source @ 35deg: 133Ba_hi
Distance: each
Start time: 11:31:47 - 30.05.2021
Stop time: 13:10:14 - 30.05.2021
file name: e127b_run0213.lmd
avrg. rate: 700Hz
dead-time: 6% |
565
|
Sun May 30 13:23:26 2021 |
Jan Glorius | Calibration | run0214 - X-calib. 90-ba 145-ba 35-am | 2021 |
Efficiency calibration in ESR after experiment
************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************
Detectors: 90 + 145 + 35
Source @ 90deg: 133Ba_lo
Source @ 145deg: 133Ba_hi
Source @ 35deg: 241Am_hi
Distance: each
Start time: 13:22:38 - 30.05.2021
Stop time: 14:28:09 - 30.05.2021
file name: e127b_run0214.lmd
avrg. rate: 1150Hz
dead-time: 8% |
568
|
Sun May 30 14:48:09 2021 |
Jan, Sophia | Calibration | run0215 - crap | 2021 |
|
566
|
Sun May 30 14:38:09 2021 |
Jan Glorius | Calibration | run0216 - X-calib. 90-ba 35-pb | 2021 |
Efficiency calibration in ESR after experiment
************************************
GEM1800 @ 90deg (xray1)
GEM3000 @ 35deg (xray2)
GLP2000 @ 145deg (xray3)
************************************
Detectors: 90 + 35
Source @ 90deg: 133Ba_lo
Source @ 35deg: 210Pb
Distance: each
Start time: 14:46:46 - 30.05.2021
Stop time: 20:11:08 - 30.05.2021
file name: e127b_run0216.lmd
avrg. rate: 330Hz
dead-time: 3% |
41
|
Sun Mar 15 09:42:40 2020 |
Jan | Calibration | run030 - Xray90 calib Am241 | |
measurement with old tennelec specAmps, do not use for efficiency!
Detector: 90
Source: Am-241
Distance: 167.5mm
Start time: 13:06:55 15.03.2020
Stop time: 13:24:57 15.03.2020
file name: run030_xxxx.lmd
avrg. rate: 650Hz
dead-time: 7% |
42
|
Sun Mar 15 13:08:38 2020 |
Jan | Calibration | run031 - Xray90 calib Ba133 hi-rate | |
measurement with old tennelec specAmps, do not use for efficiency!
Detector: 90
Source: Ba-133 strong source
Distance: 167.5mm
Start time: 13:28:42 15.03.2020
Stop time:13:38:16 15.03.2020
file name: run031_xxxx.lmd
avrg. rate: 2.7kHz
dead-time: 25% |
43
|
Sun Mar 15 13:40:11 2020 |
Jan | Calibration | run032 - Xray90 calib Ba133 | |
measurement with old tennelec specAmps, do not use for efficiency!
Detector: 90
Source: Ba-133 weak source
Distance: 167.5mm
Start time: 13:39:30 15.03.2020
Stop time: 13:50:55 15.03.2020
file name: run032_xxxx.lmd
avrg. rate: 150Hz
dead-time: 1.5% |
44
|
Sun Mar 15 13:52:05 2020 |
Jan | Calibration | run033 - Xray145 calib Am241 | |
measurement with old tennelec specAmps, do not use for efficiency!
Detector: 145
Source: Am-241
Distance: 305mm
Start time: 14:15;56 15.03.2020
Stop time: 14:40:21 15.03.2020
file name: run033_xxxx.lmd
avrg. rate: 300Hz
dead-time: 4% |
45
|
Sun Mar 15 14:41:34 2020 |
Jan | Calibration | run034 - Xray145 calib B133 high rate -SOURCE MOVED! | |
measurement with old tennelec specAmps, do not use for efficiency!
The source moved during the calibration, measurement stopped at half time.
Detector: 145
Source: B133 high rate
Distance: 305mm
Start time: 14:54:47 15.03.2020
Stop time: 15:06:53 15.03.2020
file name: run034_xxxx.lmd
avrg. rate: 1kHz
dead-time: 11% |
46
|
Sun Mar 15 15:08:44 2020 |
Jan | Calibration | run035 - Xray145 calib B133 high rate | |
measurement with old tennelec specAmps, do not use for efficiency!
Detector: 145
Source: B133 high rate
Distance: 305mm
Start time: 15:10:15 15.03.2020
Stop time: 15:31:16 15.03.2020
file name: run035_xxxx.lmd
avrg. rate: 1kHz
dead-time: 12% |
49
|
Sun Mar 15 17:14:08 2020 |
Laszlo | Calibration | run036 - Xray90 calib Am241, distance 200! | |
Detector: 90
Source: Am241
Distance: 200mm
Start time: 17:15:12 15.03.2020
Stop time: 17:32:17 15.03.2020
file name: run036_xxxx.lmd
avrg. rate: 450Hz
dead-time: 5% |
50
|
Sun Mar 15 17:33:17 2020 |
Laszlo | Calibration | run037 - Xray90 calib Am241 | |
Detector: 90
Source: Am241
Distance: 167.5mm
Start time: 17:40:00 15.03.2020
Stop time: 17:49:57 15.03.2020
file name: run037_xxxx.lmd
avrg. rate: 620Hz
dead-time: 7%
--------------------
update at 05.05.2020: the cables were probably twisted between the 90 and the 145 angle detectors. |
Attachment 1: run037_241Am_145.png
|
|
51
|
Sun Mar 15 17:50:42 2020 |
Laszlo | Calibration | run038 - Xray90 calib Ba133, high rate | |
Detector: 90
Source: Ba133 - high rate
Distance: 167.5mm
Start time: 17:53:38 15.03.2020
Stop time: 18:05:41 15.03.2020
file name: run038_xxxx.lmd
avrg. rate: 2.6kHz
dead-time: 26%
--------------------
update at 05.05.2020: the cables were probably twisted between the 90 and the 145 angle detectors. |
Attachment 1: run038_133Ba_high_145.png
|
|
52
|
Sun Mar 15 18:09:19 2020 |
Laszlo | Calibration | run039 - Xray90 calib Ba133 | |
Detector: 90
Source: Ba133
Distance: 167.5mm
Start time: 18:10:02 15.03.2020
Stop time: 18:47:36 15.03.2020
file name: run039_xxxx.lmd
avrg. rate: 150Hz
dead-time: 2% |
53
|
Sun Mar 15 18:53:42 2020 |
Laszlo | Calibration | run040 - Xray145 calib Am241 | |
Detector: 145
Source: Am241
Distance: 305mm
Start time: 18:53:16 15.03.2020
Stop time: 19:18:35 15.03.2020
file name: run040_xxxx.lmd
avrg. rate: 230Hz
dead-time: 3%
--------------------
update at 05.05.2020: the cables were probably twisted between the 90 and the 145 angle detectors. |
Attachment 1: run040_241Am_90.png
|
|
54
|
Sun Mar 15 19:22:58 2020 |
Laszlo | Calibration | run041 - Xray145 calib Ba133 - high rate | |
Detector: 145
Source: Ba133
Distance: 305mm
Start time: 19:22:34 15.03.2020
Stop time: 19:38:44 15.03.2020
file name: run041_xxxx.lmd
avrg. rate: 1kHz
dead-time: 11%
--------------------
update at 05.05.2020: the cables were probably twisted between the 90 and the 145 angle detectors. |
Attachment 1: run041_133Ba_high_90.png
|
|
55
|
Sun Mar 15 19:46:34 2020 |
Laszlo | Calibration | run042 - Xray35 calib Ba133 - high rate | |
Detector: 35
Source: Ba133, strong source
Distance: 334mm
Start time: 19:46:54 15.03.2020
Stop time: 20:20:14 15.03.2020
file name: run042_xxxx.lmd
avrg. rate: 60Hz
dead-time: 1% |
56
|
Sun Mar 15 20:26:08 2020 |
Laszlo | Calibration | run043 - Xray35 calib Am241 | |
Detector: 35
Source: Am241
Distance: 334mm
Start time: 20:25:48 15.03.2020
Stop time: 20:56:56 15.03.2020
file name: run043_xxxx.lmd
avrg. rate: 20Hz
dead-time: ~0%
--------------------
update at 05.05.2020: the cables were probably twisted between the 90 and the 145 angle detectors. |
Attachment 1: run043_241Am_35.png
|
|
58
|
Sun Mar 15 21:01:23 2020 |
Laszlo | Calibration | run044 - Xray35 calib Ba133 - a night long run | |
Detector: 35
Source: Ba133
Distance: 334mm
Start time: 21:01:26 15.03.2020
Stop time: 7:11:07 16.03.2020
file name: run044_xxxx.lmd
avrg. rate: 60Hz
dead-time: 1-2%
--------------------
update at 05.04.2020: the cables were probably twisted between the 90 and the 145 angle detectors. However, the id of the detector doesnt change, since there were no simultanious measurements (one source used with one detector a detector at a time) |
Attachment 1: run044_133Ba_high_35.png
|
|