HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te, Page 2 of 29  Not logged in ELOG logo
ID Date Author Category Subjectdown Year
  264   Tue Mar 24 00:11:35 2020 LaszloCalibrationrun116 - Xray35 calib AM241, distance 334mm 
Detector: 35
Source: 241Am
Distance: 334mm
Start time: 00:11:03 24.03.2020
Stop time:  10:01:4224.03.2020

file name: run116_xxxx.lmd
avrg. rate: 20Hz
dead-time:  0%
  263   Mon Mar 23 22:25:41 2020 LaszloCalibrationrun115 - Xray35 Ba133 d=334mm, Xray90 Lead source, d=167.5mm 
Simultaneous measurement of Xray35 and Xray90. The dead-time is <3%

Detector: 35
Source: 133Ba - strong source
Distance: 334mm

Detector: 90
Source: Lead source
Distance: 167.5mm: the lead source was measured 3mm width with the source in the middle. Therefore, I set the paper zylinder to ~169mm distance. 

Start time: 22:25:22 23.03.2020
Stop time:  00:04:49 24.03.2020

file name: run115_xxxx.lmd
avrg. rate35: 75Hz
avrg. rate90: 55Hz
dead-time:  2%

Uwe reduced the noise on the 145 degree detector as well. 

During measurement the oscilloscope remained connected!
  262   Mon Mar 23 21:47:40 2020 LaszloCalibrationrun114 - Xray145 calib Ba133, distance 305mm - LESS NOISE 
Detector: 145
Source: 133Ba - strong source
Distance: 305mm
Start time: 21:47:20 23.03.2020
Stop time:  22:04:06 23.03.2020

file name: run114_xxxx.lmd
avrg. rate: 950Hz
dead-time:  11%

Uwe reduced the noise on the 145 degree detector as well. 

During measurement the oscilloscope remained connected!
  261   Mon Mar 23 20:52:37 2020 LaszloCalibrationrun113 - Xray145 calib Am241, distance 305mm - LESS NOISE 
Detector: 145
Source: 241Am
Distance: 305mm
Start time: 20:01:43 23.03.2020
Stop time:  21:43:34

file name: run113_xxxx.lmd
avrg. rate: 220Hz
dead-time:  2.5%

Uwe reduced the noise on the 145 degree detector as well. 
  260   Mon Mar 23 19:32:53 2020 LaszloCalibrationrun112 - Xray90 calib Am241, distance 200mm - LESS NOISE 
Detector: 90
Source: 241Am
Distance: 200mm
Start time: 20:01:43 23.03.2020
Stop time:  

file name: run112_xxxx.lmd
avrg. rate: 444Hz
dead-time:  5%

Uwe reduced the noise on the 90 degree detector (see elog entry)
  267   Wed Mar 25 14:26:22 2020 LaszloCalibrationrun111 - Xray90 calib Am241, distance 167.5mm - LESS NOISE 
Detector: 90
Source: 241Am
Distance: 167.5mm
.Start time: ? 23.03.2020
Stop time:  

file name: run111_xxxx.lmd
avrg. rate: 444Hz
dead-time:  5%

Uwe reduced the noise on the 90 degree detector (see elog entry)
  259   Mon Mar 23 18:17:27 2020 LaszloCalibrationrun110 - Xray90 calib Ba133, distance 167.5mm -low rate - LESS NOISE 
Detector: 90
Source: Ba133 - weak source
Distance: 167.5mm
Start time: 18:14:50 23.03.2020
Stop time:  

file name: run110_xxxx.lmd
avrg. rate: 160Hz
dead-time:  2%

Uwe reduced the noise on the 90 degree detector (see previous elog entry)
  257   Mon Mar 23 17:40:33 2020 LaszloCalibrationrun109 - Xray90 calib Ba133, distance 167.5mm -low rate 
Detector: 90
Source: Ba133 - weak source
Distance: 167.5mm
Start time: 17:40:11 23.03.2020
Stop time:  

file name: run109_xxxx.lmd
avrg. rate: 170Hz
dead-time:  2%
  256   Mon Mar 23 17:32:42 2020 LaszloCalibrationrun108 - Xray90 calib Ba133, distance 167.5mm - high rate - deadtime too high! 
Detector: 90
Source: Ba133 - strong source
Distance: 167.5mm
Start time: 17:32:11 23.03.2020
Stop time:  17:35:34

file name: run108_xxxx.lmd
avrg. rate: 2.7kHz
dead-time:  26%
  255   Mon Mar 23 16:21:33 2020 LaszloCalibrationrun107 - Xray90 calib Am241, distance 167.5mm 
Detector: 90
Source: Am241 
Distance: 167.5mm
Start time: 15:46:49 23.03.2020
Stop time:  17:32 23.03.2020

file name: run107_xxxx.lmd
avrg. rate: 610Hz
dead-time:  7%

not Gaussian shaped peaks, after previous investigations with the scope, the scope is disconnected.

Just to mention, in order to improve deadtime the other two Xray dets are disconnected.
  254   Mon Mar 23 16:17:58 2020 LaszloCalibrationrun106 - garbage! 
run106 was clicked by accident, please delete!
  253   Mon Mar 23 15:48:15 2020 LaszloCalibrationrun105 - Xray90 calib Am241, distance 167.5mm, wrong? 
Detector: 90
Source: Am241 
Distance: 167.5mm
Start time: 15:46:49 23.03.2020
Stop time:  

file name: run105_xxxx.lmd
avrg. rate: 610Hz
dead-time:  7.5%

suspiciously not Gaussian shaped peaks --> investigation with oscilloscope
  58   Sun Mar 15 21:01:23 2020 LaszloCalibrationrun044 - Xray35 calib Ba133 - a night long run  
Detector: 35
Source: Ba133
Distance: 334mm
Start time: 21:01:26 15.03.2020
Stop time:  7:11:07 16.03.2020

file name: run044_xxxx.lmd
avrg. rate: 60Hz
dead-time:  1-2%


--------------------

update at 05.04.2020: the cables were probably twisted between the 90 and the 145 angle detectors. However, the id of the detector doesnt change, since there were no simultanious measurements (one source used with one detector a detector at a time)
  56   Sun Mar 15 20:26:08 2020 LaszloCalibrationrun043 - Xray35 calib Am241 
Detector: 35
Source: Am241
Distance: 334mm
Start time: 20:25:48 15.03.2020
Stop time:  20:56:56 15.03.2020

file name: run043_xxxx.lmd
avrg. rate: 20Hz
dead-time:  ~0%


--------------------

update at 05.05.2020: the cables were probably twisted between the 90 and the 145 angle detectors.
  55   Sun Mar 15 19:46:34 2020 LaszloCalibrationrun042 - Xray35 calib Ba133 - high rate 
Detector: 35
Source: Ba133, strong source
Distance: 334mm
Start time: 19:46:54 15.03.2020
Stop time:  20:20:14 15.03.2020

file name: run042_xxxx.lmd
avrg. rate: 60Hz
dead-time:  1%
  54   Sun Mar 15 19:22:58 2020 LaszloCalibrationrun041 - Xray145 calib Ba133 - high rate 
Detector: 145
Source: Ba133
Distance: 305mm
Start time: 19:22:34 15.03.2020
Stop time: 19:38:44 15.03.2020

file name: run041_xxxx.lmd
avrg. rate: 1kHz
dead-time:  11%


--------------------

update at 05.05.2020: the cables were probably twisted between the 90 and the 145 angle detectors.
  53   Sun Mar 15 18:53:42 2020 LaszloCalibrationrun040 - Xray145 calib Am241 
Detector: 145
Source: Am241
Distance: 305mm
Start time: 18:53:16 15.03.2020
Stop time: 19:18:35 15.03.2020

file name: run040_xxxx.lmd
avrg. rate: 230Hz
dead-time:  3%


--------------------

update at 05.05.2020: the cables were probably twisted between the 90 and the 145 angle detectors.
  52   Sun Mar 15 18:09:19 2020 LaszloCalibrationrun039 - Xray90 calib Ba133 
Detector: 90
Source: Ba133
Distance: 167.5mm
Start time: 18:10:02 15.03.2020
Stop time:  18:47:36 15.03.2020

file name: run039_xxxx.lmd
avrg. rate: 150Hz
dead-time:  2%
  51   Sun Mar 15 17:50:42 2020 LaszloCalibrationrun038 - Xray90 calib Ba133, high rate 
Detector: 90
Source: Ba133 - high rate
Distance: 167.5mm
Start time: 17:53:38 15.03.2020
Stop time: 18:05:41 15.03.2020

file name: run038_xxxx.lmd
avrg. rate: 2.6kHz
dead-time:  26%


--------------------

update at 05.05.2020: the cables were probably twisted between the 90 and the 145 angle detectors.
  50   Sun Mar 15 17:33:17 2020 LaszloCalibrationrun037 - Xray90 calib Am241 
Detector: 90
Source: Am241 
Distance: 167.5mm
Start time: 17:40:00 15.03.2020
Stop time:  17:49:57 15.03.2020

file name: run037_xxxx.lmd
avrg. rate: 620Hz
dead-time:  7%

--------------------

update at 05.05.2020: the cables were probably twisted between the 90 and the 145 angle detectors.
ELOG V3.1.5-fc6679b