HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  Nuclear hyperfine mixing in 229Th89+  Not logged in ELOG logo
Message ID: 144     Entry time: Thu Apr 7 22:18:06 2022
Author: Night shift 7.4-8.4 
Category:  
Subject: Next scans 
DAQ1: lxg1927:/data.local2/BeamTimeTh22/
DAQ2: lxg0155:/data.local2/E142/
LMD file name DAQ1:229Thor
LMD file name DAQ2:229Th

Shottky and NTCAP started at 23:53, center freq. 242.2M, 10MIQs, -30dBm
Shottky marker pos: 243.973MHz

TimeSIS-intensitycooler-HVLMD-DAQ1LMD-DAQ2scan range(nm)step width(nm)dwell time(s)comments
00:201.4e92096170065-00677810323.2-321.20.014buggy DAQ? see Elog, file 0068 empty
01:271.6e92096060069-00707811323.2-321.20.014buggy DAQ? see Elog, file 0071 empty
02:151.6e92096110072-00737812323.2-321.20.014buggy DAQ? see Elog, file 0074-0075, 7813-7814 empty
03:001.1e9?0076-00777815323.2-321.20.014buggy DAQ? see Elog and below, file 0078, 7816 empty




We were searching for a difference of a factor of twoin both DAQs. It turns out that the default value for the MBSstream server task is scale=2. In the main task this value is explicately set to scale=1. we will now (starting with filenumber: 0078) restart the 2nd DAQ with set stream_serv scale=1.

Also TDC is disabled.
From file number 0076 in DAQ2 : ADC thresholds:
unsigned short caen_V785AH_threshold_mod0 [32] =
{0x0100 , 0x0005 , 0x0004 , 0x0003 , 0x0004 , 0x0003 , 0x0005 , 0x0003 ,
0x0005 , 0x0004 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 ,
0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 ,
0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 , 0x0100 };


Before the next scan, we realized, that the laser destabilized. For the rest of the shift we tried to get laser back online (see ID:149).