Sat Apr 9 16:55:19 2022
Author:
Rodolfo
Category:
Laser
Subject:
16:15 I have changed the dye (of both oscillator and amplifier).
16:27 I have increased the pump energy (@ 532 nm) from 476 mJ to 507 mJ.
- UV output energy @ 322,8 nm: 12 mJ. (Before was at 9 mJ).
- New calibration factor is 160 for Quanta Ray monitor signal.