HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018
  Nuclear hyperfine mixing in 229Th89+  Not logged in ELOG logo
Message ID: 186     Entry time: Sat Apr 9 16:55:19 2022
Author: Rodolfo 
Category: Laser 
Subject:  
16:15 I have changed the dye (of both oscillator and amplifier).

16:27 I have increased the pump energy (@ 532 nm) from 476 mJ to 507 mJ.

- UV output energy @ 322,8 nm: 12 mJ. (Before was at 9 mJ).

- New calibration factor is 160 for Quanta Ray monitor signal.
ELOG V3.1.5-fc6679b