|   Nuclear hyperfine mixing in 229Th89+  | 
Not logged in | 
  | 
   | 
| 
Message ID: 192
    Entry time: Sun Apr 10 04:07:53 2022
 |  
| Author: | 
Patrick  | 
 | Category: | 
DAQ  | 
 | Subject: | 
go4 shows wrong frequencies and wavelengths  | 
 
  |  
 
 | 
Update (ID:194): In labview "Stop All", close labview, stop python script (ID:197), close wavemeter. Then restart wavemeter, python script, labview an mbs. 
 
Both labview and the wavemeter software are displaying the scanned wavelength correctly (Fig. 1 & 2), but go4 shows something which looks just like a freely drifting wavelength (Fig. 3). 
The Cobra set voltages are displayed correctly. | 
 | 
 | 
 
 | 
 |